Purpose: Laximeters were designed to diagnose an anterior cruciate ligament (ACL) deficient knee, but their use has now focused on providing an objective assessment of the anterior translation (AT) of an intact and ACL-reconstructed knee. In this study we report the introduction and direct comparison of an automated and computerized AT measurement device, GNRB, with the device previously established by the institute and as the current literature standard, the KT1000.
Methods: A prospective data collection was commenced upon introduction of the GNRB. The measurements of AT in each patient were performed by the same investigator with each device using 134 N applied to both knees, giving a side-to-side difference. The investigators were a sport scientist, a biomechanical engineer and a physiotherapist. Increased AT was defined as a difference > 3 mm.
Results: Three investigators performed the measurements in 122 patients, 9.8 (± 1.8) months after ACL reconstruction. Mean AT of the healthy knee was 5.7 mm with KT1000 and 4.4 mm with GNRB (p = 0.002). Mean AT of the ACL reconstructed knee was 7.0 mm with the KT1000 and 5.3 mm with the GNRB (p = 0.037). The KT1000 had a higher variance of results than the GNRB (p < 0.001). There were 25 patients with increased AT measured by KT1000 compared with 12 patients using the GNRB (p < 0.016), with only 5 on both devices.
Conclusions: GNRB has better consistency of results when compared to the KT1000. Both devices lack comparability for detecting increased AT, with the KT1000 recording a side-to-side difference of more than 3 mm in twice as many patients as the GNRB.
Level of evidence: II.
Keywords: Anterior cruciate ligament; GNRB; KT1000; Knee; Laximeter; Laximetry; Laxity.