Infections Associated with Resterilized Pacemakers and Defibrillators
- PMID: 32374963
- DOI: 10.1056/NEJMoa1813876
Infections Associated with Resterilized Pacemakers and Defibrillators
Abstract
Background: Access to pacemakers and defibrillators is problematic in places with limited resources. Resterilization and reuse of implantable cardiac devices obtained post mortem from patients in wealthier nations have been undertaken, but uncertainty around the risk of infection is a concern.
Methods: A multinational program was initiated in 1983 to provide tested and resterilized pacemakers and defibrillators to underserved nations; a prospective registry was established in 2003. Patients who received reused devices in this program were matched in a 1:3 ratio with control patients who received new devices implanted in Canada. The primary outcome was infection or device-related death, with mortality from other causes modeled as a competing risk.
Results: Resterilized devices were implanted in 1051 patients (mean [±SD] age, 63.2±18.5 years; 43.6% women) in Mexico (36.0%), the Dominican Republic (28.1%), Guatemala (26.6%), and Honduras (9.3%). Overall, 85% received pacemakers and 15% received defibrillators, with one (55.5%), two (38.8%), or three (5.7%) leads. Baseline characteristics did not differ between these patients and the 3153 matched control patients. At 2 years of follow-up, infections had occurred in 21 patients (2.0%) with reused devices and in 38 (1.2%) with new devices (hazard ratio, 1.66; 95% confidence interval, 0.97 to 2.83; P = 0.06); there were no device-related deaths. The most common implicated pathogens were Staphylococcus aureus and S. epidermidis.
Conclusions: Among patients in underserved countries who received a resterilized and reused pacemaker or defibrillator, the incidence of infection or device-related death at 2 years was 2.0%, an incidence that did not differ significantly from that seen among matched control patients with new devices in Canada.
Copyright © 2020 Massachusetts Medical Society.
Comment in
-
Resterilized pacemakers and defibrillators not associated with increased infection.Nat Rev Cardiol. 2020 Aug;17(8):454. doi: 10.1038/s41569-020-0402-z. Nat Rev Cardiol. 2020. PMID: 32467595 No abstract available.
-
Infections Associated with Resterilized Pacemakers and Defibrillators.N Engl J Med. 2020 Oct 1;383(14):1396. doi: 10.1056/NEJMc2027519. N Engl J Med. 2020. PMID: 32997925 No abstract available.
-
Infections Associated with Resterilized Pacemakers and Defibrillators.N Engl J Med. 2020 Oct 1;383(14):1396-1397. doi: 10.1056/NEJMc2027519. N Engl J Med. 2020. PMID: 32997926 No abstract available.
Similar articles
-
Assessing the safety of implantable cardioverter-defibrillator reuse-A retrospective case-control study.Pacing Clin Electrophysiol. 2019 Aug;42(8):1095-1098. doi: 10.1111/pace.13742. Epub 2019 Jun 30. Pacing Clin Electrophysiol. 2019. PMID: 31197857
-
Reuse of pacemakers: comparison of short and long-term performance.Circulation. 2013 Mar 19;127(11):1177-83. doi: 10.1161/CIRCULATIONAHA.113.001584. Epub 2013 Feb 20. Circulation. 2013. PMID: 23426104 Clinical Trial.
-
Reuse of explanted, resterilized implantable cardioverter-defibrillators: a cohort study.Ann Intern Med. 2012 Oct 16;157(8):542-8. doi: 10.7326/0003-4819-157-8-201210160-00004. Ann Intern Med. 2012. PMID: 23070487
-
Reuse of pacemakers and defibrillators in developing countries: logistical, legal, and ethical barriers and solutions.Heart Rhythm. 2010 Nov;7(11):1623-7. doi: 10.1016/j.hrthm.2010.04.027. Epub 2010 Apr 26. Heart Rhythm. 2010. PMID: 20430113 Review.
-
Pacemakers and implantable cardioverter defibrillators, unknown to chest radiography: review, complications and systematic reading.Eur J Radiol. 2015 Mar;84(3):499-508. doi: 10.1016/j.ejrad.2014.12.011. Epub 2014 Dec 22. Eur J Radiol. 2015. PMID: 25563861 Review.
Cited by
-
Complementary role of governments, non-governmental organizations, industry, and medical societies in expanding bradycardia therapy access.Eur Heart J Suppl. 2023 Dec 1;25(Suppl H):H22-H26. doi: 10.1093/eurheartjsupp/suad124. eCollection 2023 Dec. Eur Heart J Suppl. 2023. PMID: 38046889 Free PMC article.
-
Adoptive macrophage directed photodynamic therapy of multidrug-resistant bacterial infection.Nat Commun. 2023 Nov 9;14(1):7251. doi: 10.1038/s41467-023-43074-9. Nat Commun. 2023. PMID: 37945555 Free PMC article.
-
Feasibility and safety of reprocessing of intracardiac echocardiography catheters for electrophysiology procedures - a large single center experience.Cardiovasc Ultrasound. 2023 Oct 26;21(1):20. doi: 10.1186/s12947-023-00318-4. Cardiovasc Ultrasound. 2023. PMID: 37880719 Free PMC article.
-
Advancing global equity in cardiac care as cardiac implantable electronic device reuse comes of age.Heart Rhythm O2. 2022 Dec 16;3(6Part B):799-806. doi: 10.1016/j.hroo.2022.08.006. eCollection 2022 Dec. Heart Rhythm O2. 2022. PMID: 36589002 Free PMC article.
-
Safety and Efficacy of Submuscular Implantation With Resterilized Cardiac Implantable Electronic Device in Patients With Device Infection: A Retrospective Observational Study in Taiwan.Open Forum Infect Dis. 2022 Apr 10;9(5):ofac100. doi: 10.1093/ofid/ofac100. eCollection 2022 May. Open Forum Infect Dis. 2022. PMID: 35415195 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical