Modular Fluted Tapered Stems in Aseptic Oncologic Revision Total Hip Arthroplasty: A Game Changer?

J Arthroplasty. 2020 Dec;35(12):3692-3696. doi: 10.1016/j.arth.2020.06.038. Epub 2020 Jun 18.

Abstract

Background: Modular fluted tapered (MFT) stems are the most frequently used femoral component in revision total hip arthroplasties (THAs). Despite this, no data are available on how they perform in revision THA for oncologic salvage. This is a unique population, often with severe bone loss and prior radiation that extends the limits of uncemented femoral reconstruction. The aims of this study were to evaluate the implant survivorship, radiographic results, and clinical outcomes of MFT stems used for revision oncologic salvage.

Methods: We identified 17 patients treated initially with primary THA for an oncologic diagnosis (15 primary oncologic, 2 metastatic disease) who underwent subsequent femoral revision with an MFT stem. Mean age at revision was 66 years and 35% of patients were female. Mean follow-up was 4 years. Before revision, 5 of 17 had undergone local radiation.

Results: Ten-year survivorship free from aseptic loosening was 100%. The survivorship free of any reoperation was 76%. There were no femoral component fractures. Three patients were revised for recurrent instability, and 1 patient underwent irrigation and debridement for an acute infection. At most recent follow-up, no patient had radiographic evidence of progressive femoral component subsidence or failure of osteointegration. The mean Harris Hip Score improved from 29 preoperatively to 76 postoperatively (P < .0001).

Conclusion: In this series of patients with cancer, many of whom had severe bone loss and/or prior local radiation, being treated with revision THA, there were no revisions for femoral component loosening and no cases of implant fracture.

Level of evidence: III.

Keywords: aseptic loosening; femur; modular fluted tapered stems (MFT); oncologic salvage; revision total hip arthroplasty (THA).

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Female
  • Femur / diagnostic imaging
  • Femur / surgery
  • Hip Prosthesis* / adverse effects
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Treatment Outcome