Vessel-selective 4D-MR angiography using super-selective pseudo-continuous arterial spin labeling may be a useful tool for assessing brain AVM hemodynamics

Eur Radiol. 2020 Dec;30(12):6452-6463. doi: 10.1007/s00330-020-07057-4. Epub 2020 Jul 21.

Abstract

Objectives: To evaluate the usefulness of 4D-MR angiography based on super-selective pseudo-continuous ASL combined with keyhole and view-sharing (4D-S-PACK) for vessel-selective visualization and to examine the ability of this technique to visualize brain arteriovenous malformations (AVMs).

Methods: In this retrospective study, 15 patients (ten men and five women, mean age 44.0 ± 16.9 years) with brain AVMs were enrolled. All patients were imaged with 4D-PACK (non-selective), 4D-S-PACK, and digital subtraction angiography (DSA). Observers evaluated vessel selectivity, identification of feeding arteries and venous drainage patterns, visualization scores, and contrast-to-noise ratio (CNR) for each AVM component. Measurements were compared between the MR methods.

Results: Vessel selectivity was graded 4 in 43/45 (95.6%, observer 1) and 42/45 (93.3%, observer 2) territories and graded 3 in two (observer 1) and three (observer 2) territories. The sensitivity and specificity for identification of feeding arteries for both observers was 88.9% and 100% on 4D-PACK, and 100% and 100% on 4D-S-PACK, respectively. For venous drainage, the sensitivity and specificity was 100% on both methods for observer 1. The sensitivity and specificity for observer 2 was 94.4% and 83.3% on 4D-PACK, and 94.4% and 91.7% on 4D-S-PACK, respectively. The CNRs at the timepoint of 1600 ms were slightly lower in 4D-S-PACK than in 4D-PACK for all AVM components (Feeding artery, p = .02; nidus, p = .001; and draining artery, p = .02). The visualization scores for both observers were not significantly different between 4D-PACK and 4D-S-PACK for all components.

Conclusions: 4D-S-PACK could be a useful non-invasive clinical tool for assessing hemodynamics in brain AVMs.

Key points: • The 4D-MR angiography based on super-selective pseudo-continuous arterial spin labeling combined with CENTRA-keyhole and view-sharing (4D-S-PACK) enabled excellent vessel selectivity. • The 4D-S-PACK enabled the perfect identification of feeding arteries of brain arteriovenous malformation (AVM). • 4D-S-PACK could be a non-invasive clinical tool for assessing hemodynamics in brain AVMs.

Keywords: Arteriovenous malformations; Brain; Magnetic resonance angiography; Magnetic resonance imaging.

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Angiography, Digital Subtraction / methods
  • Brain / blood supply
  • Brain / diagnostic imaging
  • Brain / physiopathology
  • Child
  • Female
  • Hemodynamics / physiology*
  • Humans
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods*
  • Intracranial Arteriovenous Malformations / diagnostic imaging*
  • Intracranial Arteriovenous Malformations / physiopathology*
  • Magnetic Resonance Angiography / methods*
  • Male
  • Middle Aged
  • Retrospective Studies
  • Sensitivity and Specificity
  • Spin Labels
  • Young Adult

Substances

  • Spin Labels