Woven EndoBridge (WEB) device in the treatment of ruptured aneurysms

J Neurointerv Surg. 2021 May;13(5):443-446. doi: 10.1136/neurintsurg-2020-016405. Epub 2020 Jul 21.

Abstract

Background: Wide-necked bifurcation aneurysms (WNBAs) present unique challenges for endovascular treatment. The Woven EndoBridge (WEB) device is an intrasaccular braided device, recently approved by the FDA for treatment of WNBAs. While treatment of intracranial aneurysms with the WEB device has been shown to yield an adequate occlusion rate of 85% at 1 year, few data have been published for patients with ruptured aneurysms.

Objective: To present a multi-institutional series depicting the safety and efficacy of using the WEB device as the primary treatment modality in ruptured intracranial aneurysms.

Methods: A multi-institutional retrospective analysis was conducted, assessing patients presenting with aneurysmal subarachnoid hemorrhage treated with the WEB between January 2014 and April 2020. Baseline demographics, aneurysm characteristics, adverse events, and long-term outcomes (occlusion, re-treatment, functional status) were collected. A descriptive analysis was performed, and variables potentially associated with aneurysm recurrence or re-treatment were assessed.

Results: Forty-eight patients were included. Anterior communicating artery aneurysms were the most common (35.4%) location for treatment, followed by middle cerebral artery (20.8%) and basilar apex (16.7%). Procedural success was noted in 95.8% of patients, and clinically significant periprocedural adverse events occurred in 12.5%. After a median follow-up of 5.5 months, 54.2% of patients had follow-up angiographic imaging. Complete occlusion was seen in 61.5% of cases with adequate occlusion in 92.3%. Re-treatment was required in only 4.2% of patients during the study period. Tobacco use was significantly higher in patients with aneurysm recurrence (88.9% vs 35.7%; p=0.012). No other characteristics were associated with recurrence/re-treatment. At 30 days, 81.1% were functionally independent (modified Rankin Scale score ≤2).

Conclusion: Treatment of acutely ruptured aneurysms with the WEB device demonstrates both safety and efficacy on par with rates of conventional treatment strategies.

Keywords: aneurysm; device; embolic; hemorrhage.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Aged
  • Aneurysm, Ruptured / diagnostic imaging*
  • Aneurysm, Ruptured / surgery*
  • Cohort Studies
  • Embolization, Therapeutic / instrumentation
  • Embolization, Therapeutic / methods
  • Endovascular Procedures / instrumentation
  • Endovascular Procedures / methods*
  • Female
  • Follow-Up Studies
  • Humans
  • Intracranial Aneurysm / diagnostic imaging*
  • Intracranial Aneurysm / surgery*
  • Male
  • Middle Aged
  • Recurrence
  • Retrospective Studies
  • Self Expandable Metallic Stents*
  • Treatment Outcome