Is short-term-variation of fetal-heart-rate a better predictor of fetal acidaemia in labour? A feasibility study

PLoS One. 2020 Aug 3;15(8):e0236982. doi: 10.1371/journal.pone.0236982. eCollection 2020.

Abstract

Background: Continuous intrapartum fetal monitoring is challenging and its clinical benefits are debated. The project evaluated whether short-term-variation (STV) and other computerised fetal heart rate (FHR) parameters (baseline FHR, long-term-variation, accelerations and decelerations) predicted acidaemia at birth. The aims of the study were to assess the changes in FHR pattern during labour and determine the feasibility of undertaking a definitive trial by reporting the practicalities of using the monitoring device, participant recruitment, data collection and staff training.

Methods: 200 high-risk women carrying a term singleton, non-anomalous fetus, requiring continuous FHR monitoring in labour were consented to participate from the Jessop Wing maternity unit, Sheffield, UK. The trans-abdominal fetal ECG monitor was placed as per clinical protocol. During the monitoring session, clinicians were blinded to the computerised FHR parameters. We analysed the last hour of the FHR and its ability to predict umbilical arterial blood pH <7.20 using receiver operator characteristics (ROC) curves.

Results: Of 200 women, 137 cases were excluded as either the monitor did not work from the onset of labour (n = 30), clinical staff did not return or used the monitor on another patient (n = 37), umbilical cord blood not obtained (n = 25), FHR data not recorded within an hour of birth (n = 34) and other reasons (n = 11). In 63 cases included in the final analysis, the computer-derived FHR parameters did not show significant correlation with umbilical artery cord pH <7.20. Labour was associated with a significant increase in short and long term variation of FHR and number of deceleration (P<0.001). However, baseline FHR decreased significantly before delivery (P<0.001).

Conclusions: The project encountered a number of challenges, with learning points crucial to informing the design of a large study to evaluate the potential place of intrapartum computerised FHR parameters, using abdominal fetal ECG monitor before its clinical utility and more widespread adoption can be ascertained.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acidosis / physiopathology
  • Adult
  • Cardiotocography
  • Electrocardiography
  • Electroencephalography / instrumentation*
  • Feasibility Studies
  • Female
  • Fetal Blood
  • Fetal Diseases / physiopathology
  • Fetal Monitoring / methods*
  • Fetus / physiopathology
  • Heart Rate, Fetal / physiology*
  • Humans
  • Hydrogen-Ion Concentration
  • Labor, Obstetric
  • Pregnancy