Fully Automated Lab-On-A-Disc Platform for Loop-Mediated Isothermal Amplification Using Micro-Carbon-Activated Cell Lysis

Sensors (Basel). 2020 Aug 22;20(17):4746. doi: 10.3390/s20174746.

Abstract

Fast and fully automated deoxyribonucleic acid (DNA) amplification methods are of interest in the research on lab-on-a-disc (LOD) platforms because of their full compatibility with the spin-column mechanism using centrifugal force. However, the standard procedures followed in DNA amplification require accurate noncontact temperature control as well as cell lysis at a low temperature to prevent damage to the LOD platform. This requirement makes it challenging to achieve full automation of DNA amplification on an LOD. In this paper, a fully automated LOD capable of performing cell lysis and amplification on a single compact disc of DNA samples is proposed. The proposed system uses micro-carbon to heat DNA samples without damaging the LOD as well as a noncontact heating system and an infrared camera sensor to remotely measure the real temperature of the amplification chamber. Compared with conventional DNA amplification systems, the proposed system has the advantage of full automation of the LOD platform. Experimental results demonstrated that the proposed system offers a stable heating method for DNA amplification and cell lysis.

Keywords: DNA amplification; cell lysis; infrared camera; lab-on-a-disc; micro-carbon.

MeSH terms

  • Automation
  • Carbon*
  • Lab-On-A-Chip Devices*
  • Molecular Diagnostic Techniques*
  • Nucleic Acid Amplification Techniques*

Substances

  • Carbon

Supplementary concepts

  • LAMP assay