Early aortic dissection after aortic valve replacement with a Perceval sutureless bioprosthesis

Indian J Thorac Cardiovasc Surg. 2019 Jul;35(3):510-512. doi: 10.1007/s12055-018-0675-6. Epub 2018 Apr 14.

Abstract

We report a patient who developed ascending aorta dissection after aortic valve replacement with a Perceval sutureless pericardial bioprosthesis, a previously not reported complication with this device. At reoperation, the intimal tear was located at the aortotomy suture line; the tips of the nitinol stent were far from the aortic lesion which therefore was not attributable to the presence of this specific prosthetic model. Nevertheless, a previous report of aortic dissection due to erosion of the aortic wall by the strut of a transcatheter prosthesis may raise the suspicion that such complication could also occur with sutureless prostheses of similar design.

Keywords: Aortic dissection; Aortic valve replacement; Sutureless bioprostheses.

Publication types

  • Case Reports