Multiwavelength confocal laser scanning microscopy of the cornea

Biomed Opt Express. 2020 Sep 18;11(10):5689-5700. doi: 10.1364/BOE.397615. eCollection 2020 Oct 1.

Abstract

Confocal reflectance microscopy has demonstrated the ability to produce in vivo images of corneal tissue with sufficient cellular resolution to diagnose a broad range of corneal conditions. To investigate the spectral behavior of corneal reflectance imaging, a modified laser ophthalmoscope was used. Imaging was performed in vivo on a human cornea as well as ex vivo on porcine and lamb corneae. Various corneal layers were imaged at the wavelengths 488 nm, 518 nm, and 815 nm and compared regarding image quality and differences in the depicted structures. Besides the wavelength- and depth-dependent scattering background, which impairs the image quality, a varying spectral reflectance of certain structures could be observed. Based on the obtained results, this paper emphasizes the importance of choosing the appropriate light source for corneal imaging. For the examination of the epithelial layers and the endothelium, shorter wavelengths should be preferred. In the remaining layers, longer wavelength light has the advantage of less scattering loss and a potentially higher subject compliance.