Subject validation of reusable N95 stop-gap filtering facepiece respirators in COVID-19 pandemic

PLoS One. 2020 Nov 13;15(11):e0242304. doi: 10.1371/journal.pone.0242304. eCollection 2020.

Abstract

Introduction: The COVID-19 pandemic has unveiled widespread shortages of personal protective equipment including N95 respirators. Several centers are developing reusable stop-gap respirators as alternatives to disposable N95 respirators during public health emergencies, using techniques such as 3D-printing, silicone moulding and plastic extrusion. Effective sealing of the mask, combined with respiratory filters should achieve 95% or greater filtration of particles less than 1um. Quantitative fit-testing (QNFT) data from these stop-gap devices has not been published to date. Our team developed one such device, the "SSM", and evaluated it using QNFT.

Methods: Device prototypes were iteratively evaluated for comfort, breathability and communication, by team members wearing them for 15-30min. The fit and seal were assessed by positive and negative pressure user seal checks. The final design was then formally tested by QNFT, according to CSA standard Z94.4-18 in 40 volunteer healthcare providers. An overall fit-factor >100 is the passing threshold. Volunteers were also tested by QNFT on disposable N95 masks which had passed qualitative fit testing (QLFT) by institutional Occupational Health and Safety Department.

Results: The SSM scored 3.5/5 and 4/5 for comfort and breathability. The median overall harmonic mean fit-factors of disposable N95 and SSM were 137.9 and 6316.7 respectively. SSM scored significantly higher than disposable respirators in fit-test runs and overall fit-factors (p <0.0001). Overall passing rates in disposable and SSM respirators on QNFT were 65% and 100%. During dynamic runs, passing rates in disposable and SSM respirators were 68.1% and 99.4%; harmonic means were 73.7 and 1643.

Conclusions: We present the design and validation of a reusable N95 stop-gap filtering facepiece respirator that can match existent commercial respirators. This sets a precedence for adoption of novel stop-gap N95 respirators in emergency situations.

Publication types

  • Validation Study

MeSH terms

  • Adult
  • Betacoronavirus
  • COVID-19
  • Coronavirus Infections / prevention & control*
  • Equipment Design*
  • Equipment Failure Analysis
  • Equipment Reuse
  • Female
  • Filtration
  • Humans
  • Male
  • Middle Aged
  • Pandemics / prevention & control*
  • Pneumonia, Viral / prevention & control*
  • Respiratory Protective Devices / standards*
  • SARS-CoV-2

Grants and funding

The authors received no specific funding for this work.