Abstract
Background: Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a standardized fluoroscopy-guided implantation technique.
Methods: For this cadaver study, SNM electrodes were implanted bilaterally in 5 lower body specimens. The lower edge of the sacroiliac joint and the medial edge of the sacral foramina were marked using fluoroscopy to draw an 'H' with the crossing points identifying S3. After electrode placement the pelvis was dissected to describe the exact position of the SNM electrodes.
Results: The electrodes were inserted at an angle with a median degree measure of 60° (range 50-65°) to the skin, with a median distance of 9 mm (range 0-13 mm) from the S3 marking. All electrodes entered the third sacral foramen. The median distance of the electrodes to the sacral nerve was 0 mm (range 0-3 mm) for the most proximal, 0.5 mm (range 0-5 mm) for the second, 2.25 mm (range 0-11 mm) for the third and 1.75 mm (range 0-16 mm) for the most distant electrode. There was neither a significant difference in the proximity of the electrodes to the nerve between the right and left side (proximal to distal electrode: p = 0.18, p = 0.16, p = 0.07, p = 0.07) nor between male and female cadavers (p = 0.25, p = 0.21, p = 0.66, p = 0.66).
Conclusions: A standardized fluoroscopy-guided implantation technique enables a close contact between electrode and nerve. This can potentially result in an improved clinical outcome.
Keywords: Anatomical landmarks; Electrode localization; Faecal incontinence; Lead placement; Sacral neuromodulation; Treatment success.
Conflict of interest statement
The authors do not have any personal, academic or financial conflicts of interest.
Figures
Similar articles
-
Comparison of surgical techniques for optimal lead placement in sacral neuromodulation: a cadaver study.Tech Coloproctol. 2022 Sep;26(9):707-712. doi: 10.1007/s10151-022-02632-x. Epub 2022 May 28. Tech Coloproctol. 2022. PMID: 35633425 Free PMC article.
-
Sacral Neuromodulation: Standardized Electrode Placement Technique.Neuromodulation. 2017 Dec;20(8):816-824. doi: 10.1111/ner.12695. Epub 2017 Oct 4. Neuromodulation. 2017. PMID: 28975677
-
Optimizing electrode implantation in sacral nerve stimulation--an anatomical cadaver study controlled by a laparoscopic camera.Int J Colorectal Dis. 2008 Jan;23(1):85-91. doi: 10.1007/s00384-007-0367-y. Epub 2007 Aug 18. Int J Colorectal Dis. 2008. PMID: 17704926
-
Sacral neuromodulation in patients with congenital faecal incontinence. Special issues and review of the literature.Tech Coloproctol. 2018 Feb;22(2):89-95. doi: 10.1007/s10151-017-1742-5. Epub 2018 Jan 16. Tech Coloproctol. 2018. PMID: 29340832 Review.
-
Strategies for Difficult Fluoroscopic Landmarking During Sacral Neuromodulation Lead Placement.Urology. 2023 Apr;174:218-220. doi: 10.1016/j.urology.2022.12.029. Epub 2023 Jan 11. Urology. 2023. PMID: 36638971 Review.
Cited by
-
Technical and functional outcome after sacral neuromodulation using the "H" technique.Wien Klin Wochenschr. 2023 Aug;135(15-16):399-405. doi: 10.1007/s00508-022-02115-x. Epub 2022 Dec 6. Wien Klin Wochenschr. 2023. PMID: 36472709 Free PMC article.
-
Innovative training modality for sacral neuromodulation (SNM): Patient-specific computerized tomography (CT) reconstructed 3D-printed training system: ICS School of Modern Technology novel training modality.Neurourol Urodyn. 2023 Jan;42(1):297-302. doi: 10.1002/nau.25083. Epub 2022 Nov 2. Neurourol Urodyn. 2023. PMID: 36321797 Free PMC article.
-
Comparison of surgical techniques for optimal lead placement in sacral neuromodulation: a cadaver study.Tech Coloproctol. 2022 Sep;26(9):707-712. doi: 10.1007/s10151-022-02632-x. Epub 2022 May 28. Tech Coloproctol. 2022. PMID: 35633425 Free PMC article.
References
-
- Duelund-Jakobsen J, Lundby L, Lehur PA, Wyart V, Laurberg S, Buntzen S. Is the efficacy of sacral nerve stimulation for faecal incontinence dependent on the number of active electrode poles achieved during permanent lead insertion? Colorectal Dis. 2016;18(11):O414–O419. doi: 10.1111/codi.13509. - DOI - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources