Minimizing echo and repetition times in magnetic resonance imaging using a double half-echo k-space acquisition and low-rank reconstruction

NMR Biomed. 2021 Apr;34(4):e4458. doi: 10.1002/nbm.4458. Epub 2020 Dec 9.

Abstract

Sampling k-space asymmetrically (ie, partial Fourier sampling) in the readout direction is a common way to reduce the echo time (TE) during magnetic resonance image acquisitions. This technique requires overlap around the center of k-space to provide a calibration region for reconstruction, which limits the minimum fractional echo to ~60% before artifacts are observed. The present study describes a method for reconstructing images from exact half echoes using two separate acquisitions with reversed readout polarity, effectively providing a full line of k-space without additional data around central k-space. This approach can benefit sequences or applications that prioritize short TE, short inter-echo spacing or short repetition time. An example of the latter is demonstrated to reduce banding artifacts in balanced steady-state free precession.

Keywords: low rank; partial Fourier.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*