Detailed study of the distribution of activation inside the magnet coils of a compact PET cyclotron

Appl Radiat Isot. 2021 Feb:168:109446. doi: 10.1016/j.apradiso.2020.109446. Epub 2020 Sep 28.

Abstract

We determined the distribution of activation products inside the magnet coils of a medical cyclotron that has been operational for fifteen years. Besides FLUKA, we based our approach on new software tools (RAW and ActiWiz) developed for high-energy accelerators at CERN. A combined analysis of measurements on the coils with Monte-Carlo simulations resulted in a detailed three-dimensional radiological characterisation of the coils. Our results provide the required information for the radiation protection expert to identify the appropriate waste elimination scheme.

Keywords: Activation; Cyclotron; Decommissioning; Gamma spectrometry; Magnet coil; Monte Carlo simulation; Neutron spectrum; Positron emission tomography.