Mid-Term Survivorship, Performance, and Predictors of Outcome in Primary Total Hip Arthroplasty With a Porous Tantalum Femoral Prosthesis

J Arthroplasty. 2021 May;36(5):1688-1694. doi: 10.1016/j.arth.2020.12.001. Epub 2020 Dec 16.

Abstract

Background: Femoral stem subsidence is a known cause of early implant failure, increasing the risk for aseptic loosening and periprosthetic fracture. Overall survivorship and subsidence in a novel porous tantalum-coated femoral prosthesis have not been well-studied.

Methods: Consecutive patients undergoing primary total hip arthroplasty with a porous tantalum-coated femoral prosthesis between January 2008 and January 2015 with minimum 5-year follow-up were included. Clinical and radiographic data were obtained from hospital and office records. Multivariate logistic regression analyses were used to determine predictors of subsidence and clinical outcomes. Kaplan-Meier survivorship curves were performed to illustrate primary failure endpoints of (1) all-cause revision and (2) femoral prosthesis revision.

Results: A total of 398 patients with a mean (±standard deviation) age of 61.0 ± 11.5 years, body mass index (BMI) 32.8 ± 8.0 kg/m2, and follow-up of 6.9 (range 5.0-11.2 years) were included. Survivorship at 5 years was 94.9% for all-cause revision and 98.0% for femoral component revision. Average subsidence was 1.8 ± 1.3 mm (range 0-15.5), with 8.1% patients experiencing subsidence >5 mm. Statistically significant mean improvements were observed at latest follow-up in the Hip Disability and Osteoarthritis Outcome Score Joint Replacement (40.6 ± 11.5 vs 85.2 ± 10.1, P < .001), Harris Hip Score (38.0 ± 12.0 vs 79.5 ± 12.8, P < .001), and hip flexion (92.8° ± 15.3° vs 103.3° ± 10.3°, P < .001). Multivariate logistic regression analyses revealed that greater BMI (odds ratio [OR] 1.17, P < .001), non-white/Caucasian race (OR 2.0, P = .036), and female gender (OR 2.4; P = .005) conferred a higher likelihood of subsidence >3 mm. BMI was a statistically significant and independent predictor of subsidence >5 mm (OR 1.25, P < .001) and subsidence >7 mm (OR 1.25, P < .001).

Conclusion: The trabecular metal taper femoral prosthesis conferred excellent clinical outcome improvement and survivorship. Increasing BMI was independently associated with an increased risk of subsidence in these patients and caution is recommended in utilizing this implant in obese, morbidly obese, and super morbidly obese populations.

Keywords: aseptic loosening; outcomes; porous tantalum; subsidence; total hip arthroplasty.

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Child
  • Child, Preschool
  • Female
  • Follow-Up Studies
  • Hip Prosthesis*
  • Humans
  • Obesity, Morbid*
  • Porosity
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Survivorship
  • Tantalum
  • Treatment Outcome

Substances

  • Tantalum