Background: This study aimed to compare the failure load of suture anchors used in rotator cuff repair between normal and osteoporotic bone models.
Methods: A total of 16 anchors made from metal (TwinFix Ti 5.0 or 6.5 mm, Corkscrew FT 4.5, 5.5, or 6.5 mm), polyether ether ketone (HEALICOIL PK [HC-PK] 4.5 or 5.5 mm, SwiveLock PK 4.75 or 5.5 mm), or bioabsorbable material (HEALICOIL RG [HC-RG] 4.75 or 5.5 mm, Corkscrew Bio 4.75, 5.5, or 6.5 mm, SwiveLock BC 4.75 or 5.5 mm) were included. Moreover, 10- and 5-pounds per cubic foot (pcf) Sawbone® models were set as normal and osteoporotic cancellous bone models, respectively. Pullout testing was performed in parallel to the insertion axis at a displacement rate of 12.5 mm/s using a universal testing machine. To evaluate the change in failure load between the two Sawbone® models with different densities, the remaining failure load ratio (RFLR) was defined as the ratio of the failure load in 10 pcf to that in 5 pcf.
Results: In the 10-pcf Sawbone®, TwinFix Ti 6.5 mm showed the highest mean failure load (304.0 ± 15.2 N). In the 5-pcf Sawbone® model, HC-PK 5.5 mm showed the highest failure load (146.3 ± 5.8 N). Among anchors with the same diameter, HC-PK and HC-RG showed a significantly higher failure load than other anchors in the 10- and 5-pcf Sawbone® models. HC-PK 5.5 mm (62.1%) and HC-PK 4.5 mm (51.1%) have the highest RFLR among anchors with the same diameter.
Conclusions: HC-PK and HC-RG showed higher failure load than the other anchors in both normal and osteoporotic bone models, except for TwinFix Ti 6.5 mm in the 10-pcf Sawbone® model. Based on our results, bioabsorbable anchors had sufficient failure load for rotator cuff repair in addition to bioabsorbability.
Copyright © 2021 The Japanese Orthopaedic Association. Published by Elsevier B.V. All rights reserved.