Objective measurement of adherence to wearing foot orthoses using an embedded temperature sensor

Med Eng Phys. 2021 Feb:88:19-24. doi: 10.1016/j.medengphy.2020.12.002. Epub 2020 Dec 13.

Abstract

The objective of this study was to evaluate the validity of a temperature sensor for the measurement of adherence to wearing foot orthoses. Ten participants were provided with foot orthoses containing an embedded temperature sensor and wore the orthoses for a randomly-determined duration over a five-day period. Sensor-detected wear time was compared to a reference standard (objectively measured wear time using a smart-phone application). Ambient temperature and physical activity were recorded with a temperature gauge and wearable activity tracker, respectively. A simple peak detection algorithm which identified the largest one-minute changes in sensor temperature provided highly accurate wear time values (r = 0.999, coefficient of variation=0.2%). Ambient temperature and physical activity did not significantly influence temperature sensor scores. These findings demonstrate that the temperature sensor provides accurate foot orthosis wear time data and may therefore be a useful tool for documenting adherence in clinical practice and intervention studies.

Keywords: Adherence; Footwear; Orthosis; Temperature.