A Novel Temperature-Controlled Radiofrequency Catheter Ablation System Used to Treat Patients With Paroxysmal Atrial Fibrillation

JACC Clin Electrophysiol. 2021 Mar;7(3):352-363. doi: 10.1016/j.jacep.2020.11.009. Epub 2021 Jan 27.

Abstract

Objectives: DIAMOND-AF (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation) was a prospective, multicenter, noninferiority, randomized trial that compared the safety and effectiveness of the DTA system versus those of a force-sensing RF ablation system (control) for the treatment of patients with drug-refractory, recurrent, symptomatic paroxysmal atrial fibrillation (AF).

Background: Irrigated radiofrequency (RF) ablation catheters lose tissue temperature acuity, which is vital in assessing lesion formation. DiamondTemp Ablation (DTA) was designed to re-establish accurate tissue temperature measurements during ablation.

Methods: A total of 482 patients with paroxysmal AF were randomized (239 DTA, 243 control) to undergo pulmonary vein isolation and were followed up at 23 sites. Patients were screened for disease progression, cardiac characteristics, and prior interventions. Primary endpoints were effectiveness (freedom from atrial arrhythmia recurrence) and safety (composite of procedure- and device-related serious adverse events).

Results: The primary safety event rate was 3.3% in the DTA group versus 6.6% in the control group (p < 0.001 vs. 6.5% noninferiority margin). Primary effectiveness was met in 79.1% of DTA subjects and 75.7% of control subjects (p < 0.001 vs. -12.5% noninferiority margin). Secondary endpoint analysis found that off-drug effectiveness favored DTA compared with the control (142 [59.4%] vs. 120 [49.4%], respectively; p = 0.03). Total RF time and individual RF ablation duration were significantly shorter with less saline infused through the DTA catheter (p < 0.001). Both arms saw clinically meaningful improvements in quality of life at 12 months.

Conclusions: Safety and efficacy of the DTA system proved noninferior to force-sensing RF ablation in a paroxysmal AF population. Efficiencies were observed using DTA with shorter total RF times, individual RF ablation durations, and less saline infusion. (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation; NCT03334630).

Keywords: DiamondTemp; atrial fibrillation; catheter ablation; contact force; radiofrequency.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Atrial Fibrillation* / surgery
  • Catheter Ablation* / adverse effects
  • Catheters
  • Humans
  • Prospective Studies
  • Quality of Life
  • Temperature
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT03334630