An Emerging Method to Assess Tear Film Spread and Dynamics as Possible Tear Film Homeostasis Markers

Curr Eye Res. 2021 Sep;46(9):1291-1298. doi: 10.1080/02713683.2021.1887270. Epub 2021 Feb 16.

Abstract

Purpose: This study aims to assess the performance of an analysis method to measure in vivo the movement speed of tear film particles post-blink as a measure of tear film spreading.Materials and methods: Ocular surface parameters and the recording of tear film particles' spreading post-blink were assessed in eighty-one healthy volunteers (43.7 ± 27.0 years) using Keratograph 5 M. The developed software automatically decomposed the video into frames to manually track particles' position for 1.75 seconds after a blink. The following tear film-dynamic metrics were automatically calculated: mean, median, maximum, and minimum particles' speed at different times after blinking and time for particle speed to decrease to <1.20 mm/second. Repeatability of each tear film-dynamic metric and its correlations with ocular surface signs and symptoms were analyzed. Binomial logistic regression was performed to assess the predictability of new metrics to ocular parameters.Results: Repeatability tended to be lower just after blinking (variability of 12.24%), whereas the metrics from 0.5 s onwards had acceptable repeatability (variability below 10%). Tear film-dynamic metrics correlated positively with Non-Invasive Break-Up Time (NIKBUT) while negatively with meibomian gland drop-out. Binomial logistic regression analysis revealed that tear film-dynamic metrics were able to predict NIKBUT. Nevertheless, no statistically significant association was found with gland drop-out. This means that higher particle speed is related to larger NIKBUT. The metric "time for particle speed to decrease to <1.20 mm/second" can be considered the best metric to assess the quality of the tear film, since it was more strongly correlated with NIKBUT (r = 0.42, p = .004), it was more strongly associated in the binomial logistic regression analysis with NIKBUT and showed good repeatability (variability = 5.49%).Conclusions: Tear film-dynamic metrics are emerging homeostasis parameters for assessing indirectly the tear film quality in natural conditions with acceptable repeatability.

Keywords: Dry Eye; objective medical image analysis; tear film; tear film homeostasis; tear hydrodynamics.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Blinking / physiology*
  • Dry Eye Syndromes / diagnosis*
  • Dry Eye Syndromes / metabolism
  • Dry Eye Syndromes / physiopathology
  • Female
  • Healthy Volunteers
  • Homeostasis / physiology*
  • Humans
  • Male
  • Meibomian Glands / metabolism*
  • Meibomian Glands / physiopathology
  • Middle Aged
  • Tears / physiology*
  • Young Adult