Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices With Abandoned Leads
- PMID: 33595595
- PMCID: PMC7890450
- DOI: 10.1001/jamacardio.2020.7572
Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices With Abandoned Leads
Abstract
Importance: Magnetic resonance imaging (MRI) is the modality of choice for many conditions. Conditional devices and novel protocols for imaging patients with legacy cardiac implantable electronic devices (CIEDs) have increased access to MRI in patients with devices. However, the presence of abandoned leads remains an absolute contraindication.
Objective: To assess if the performance of an MRI in the presence of an abandoned CIED lead is safe and whether there are deleterious effects on concomitant active CIED leads.
Design, setting, and participants: This cohort study included consecutive CIED recipients undergoing 1.5-T MRI with at least 1 abandoned lead between January 2013 and June 2020. MRI scans were performed at the Hospital of the University of Pennsylvania. No patients were excluded.
Exposures: CIEDs were reprogrammed based on patient-specific pacing needs. Electrocardiography telemetry and pulse oximetry were monitored continuously, and live contact with the patient throughout the scan via visual and voice contact was performed if possible. After completion of the MRI, CIED evaluation was repeated and programming returned to baseline or to a clinically appropriate setting.
Main outcomes and measures: Variation in pre- and post-MRI capture threshold of 50% or more, ventricular sensing 40% or more, and lead impedance 30% or more, as well as clinical sequelae such as pain and sustained tachyarrhythmia were considered significant. Long-term follow-up lead-related data were analyzed if available.
Results: A total of 139 consecutive patients (110 men [79%]) with a mean (SD) age of 65.6 (13.4) years underwent 200 MRIs of various anatomic regions including the thorax. Repeat examinations were common with a maximum of 16 examinations for 1 patient. There was a total of 243 abandoned leads with a mean (SD) of 1.22 (0.45) per patient. The mean (SD) number of active leads was 2.04 (0.78) and 64 patients (46%) were pacemaker dependent. A transmit-receive radiofrequency coil was used in 41 patients (20.5%), all undergoing MRI of the brain. There were no abnormal vital signs or sustained tachyarrhythmias. No changes in battery voltage, power-on reset events, or changes of pacing rate were noted. CIED parameter changes including decreased right atrial sensing in 4 patients and decreased left ventricular R-wave amplitude in 1 patient were transiently noted. One patient with an abandoned subcutaneous array experienced sternal heating that subsided on premature cessation of the study.
Conclusions and relevance: The risk of MRI in patients with abandoned CIED leads was low in this large observational study, including patients who underwent examination of the thorax. The growing aggregate of data questions the absolute contraindication for MRI in patients with abandoned CIED leads.
Conflict of interest statement
Figures
Comment in
-
Removing Obstacles to Magnetic Resonance Imaging for Patients With a Pacemaker or a Defibrillator.JAMA Cardiol. 2021 May 1;6(5):556-557. doi: 10.1001/jamacardio.2020.7593. JAMA Cardiol. 2021. PMID: 33595594 No abstract available.
-
Beyond the AJR: "Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices With Abandoned Leads".AJR Am J Roentgenol. 2021 Nov;217(5):1247. doi: 10.2214/AJR.21.26005. Epub 2021 Apr 21. AJR Am J Roentgenol. 2021. PMID: 33881901 No abstract available.
Similar articles
-
MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads.Radiol Cardiothorac Imaging. 2024 Jun;6(3):e230303. doi: 10.1148/ryct.230303. Radiol Cardiothorac Imaging. 2024. PMID: 38869431 Free PMC article.
-
Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols.Europace. 2017 May 1;19(5):812-817. doi: 10.1093/europace/euw122. Europace. 2017. PMID: 27256419
-
Magnetic resonance imaging safety in patients with abandoned or functioning epicardial pacing leads.Eur Radiol. 2022 Jun;32(6):3830-3838. doi: 10.1007/s00330-021-08469-6. Epub 2022 Jan 6. Eur Radiol. 2022. PMID: 34989847
-
State-of-the-art approach towards magnetic resonance imaging of the nervous system structures in patients with cardiac implantable electronic devices.Neurol Neurochir Pol. 2018 Nov-Dec;52(6):652-656. doi: 10.1016/j.pjnns.2018.07.002. Epub 2018 Jul 20. Neurol Neurochir Pol. 2018. PMID: 30061002 Review.
-
Performing Cardiac Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices: A Contemporary Review.Can J Cardiol. 2018 Dec;34(12):1682-1686. doi: 10.1016/j.cjca.2018.09.009. Epub 2018 Sep 29. Can J Cardiol. 2018. PMID: 30527159 Review.
Cited by
-
MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads.Radiol Cardiothorac Imaging. 2024 Jun;6(3):e230303. doi: 10.1148/ryct.230303. Radiol Cardiothorac Imaging. 2024. PMID: 38869431 Free PMC article.
-
Pre- and post-procedural cardiac imaging (computed tomography and magnetic resonance imaging) in electrophysiology: a clinical consensus statement of the European Heart Rhythm Association and European Association of Cardiovascular Imaging of the European Society of Cardiology.Europace. 2024 May 2;26(5):euae108. doi: 10.1093/europace/euae108. Europace. 2024. PMID: 38743765 Free PMC article. Review.
-
How we do it: Cardiac implantable devices are not a contraindication to MRI: time for a paradigm shift.Pediatr Radiol. 2024 May;54(6):863-875. doi: 10.1007/s00247-024-05902-y. Epub 2024 Mar 15. Pediatr Radiol. 2024. PMID: 38488925
-
Stepping back for good reasons: a reappraisal of the DF-1 connector for defibrillator leads.Europace. 2024 Mar 1;26(3):euae057. doi: 10.1093/europace/euae057. Europace. 2024. PMID: 38412340 Free PMC article.
-
SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device.J Cardiovasc Magn Reson. 2024 Summer;26(1):100995. doi: 10.1016/j.jocmr.2024.100995. Epub 2024 Jan 12. J Cardiovasc Magn Reson. 2024. PMID: 38219955 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical