Concordance of CSF measures of Alzheimer's pathology with amyloid PET status in a preclinical cohort: A comparison of Lumipulse and established immunoassays

Alzheimers Dement (Amst). 2021 Feb 6;13(1):e12131. doi: 10.1002/dad2.12131. eCollection 2021.

Abstract

Introduction: We assessed the concordance of cerebrospinal fluid (CSF) amyloid beta (Aβ) and tau measured on the fully automated Lumipulse platform with pre-symptomatic Alzheimer's disease (AD) pathology on amyloid positron emission tomography (PET).

Methods: In 72 individuals from the Insight 46 study, CSF Aβ40, Aβ42, total tau (t-tau), and phosphorylated tau at site 181 (p-tau181) were measured using Lumipulse, INNOTEST, and Meso Scale Discovery (MSD) assays and inter-platform Pearson correlations derived. Lumipulse Aβ42 measures were adjusted to incorporate standardization to certified reference materials. Logistic regressions and receiver operating characteristics analysis generated CSF cut-points optimizing concordance with 18F-florbetapir amyloid PET status (n = 63).

Results: Measurements of CSF Aβ, p-tau181, and their ratios correlated well across platforms (r 0.84 to 0.94, P < .0001); those of t-tau and t-tau/Aβ42 correlated moderately (r 0.57 to 0.79, P < .0001). The best concordance with amyloid PET (100% sensitivity and 94% specificity) was afforded by cut-points of 0.075 for Lumipulse Aβ42/Aβ40, 0.087 for MSD Aβ42/Aβ40 and 17.3 for Lumipulse Aβ42/p-tau181.

Discussion: The Lumipulse platform provides comparable sensitivity and specificity to established CSF immunoassays in identifying pre-symptomatic AD pathology.

Keywords: Lumipulse; amyloid; cerebrospinal fluid; positron emission tomography; tau.