Purpose: To evaluate bridging stent geometry in patients who underwent branched endovascular aortic repair (B-EVAR) and to correlate the outcomes with intrinsic bridging stent characteristics aiming to identify the stent(s) that guarantees the best performance.
Methods: Pre-operative and post-operative computed tomography images of all patients undergoing B-EVAR between September 2016 and April 2019 were retrospectively analyzed. Following geometrical features were measured: target vessel take-off angle (TOA); longitudinal stent shortening; shape index (SI), intended as ratio between minimum and maximum diameter of the lumen cross sections, averaged on three segments: zone 1 (proximal stented zone), zone 2 (intermediate), and zone 3 (distal).
Results: Thirty-eight branches (8 right (RRA) and 8 left renal arteries (LRA), 11 superior mesenteric arteries (SMA), 11 celiac trunks (CTR)) were treated. Fluency (Bard Peripheral Vascular), COVERA (Bard Peripheral Vascular), and VBX (WLGore&Assoc) stent-grafts were implanted in 10, 12, and 16 branches, respectively. Pre-operative TOA was more acute in RRA and LRA when compared to CTR and SMA, and straightened in post-operative configuration (109.86 ± 28.65° to 150.27 ± 21.0°; P < 0.001). Comparable values of SI among the stent types were found in zone 1 (P = 0.08), whereas higher SI in VBX group was detected in zones 2 (P < 0.001) and 3 (P < 0.001). The VBX group was also the most affected by stent shortening (11.12 ± 5.65%; P = 0.001).
Conclusion: Our early experience showed that the VBX stent offers greater stent circularity than the other devices even if a greater shortening has been observed drawing attention with regards to the decision of the nominal stent length.
Keywords: B-EVAR; Balloon-expanding stent-graft; Bridging stent; Geometric analysis; Self-expanding stent-graft; Thoraco-abdominal aortic aneurysm.