A robotically assisted 3D printed quality assurance lung phantom for Calypso
- PMID: 33657537
- DOI: 10.1088/1361-6560/abebaa
A robotically assisted 3D printed quality assurance lung phantom for Calypso
Abstract
Purpose. Radiation dose delivered to targets located near the upper-abdomen or in the thorax are significantly affected by respiratory-motion. Relatively large-margins are commonly added to compensate for this motion, limiting radiation-dose-escalation. Internal-surrogates of target motion, such as a radiofrequency (RF) tracking system, i.e. Calypso®System, are used to overcome this challenge and improve normal-tissue sparing. RF tracking systems consist of implanting transponders in the vicinity of the tumor to be tracked using radiofrequency-waves. Unfortunately, although the manufacture provides a universal quality-assurance (QA) phantom, QA-phantoms specifically for lung-applications are limited, warranting the development of alternative solutions to fulfil the tests mandated by AAPM's TG142. Accordingly, our objective was to design and develop a motion-phantom to evaluate Calypso for lung-applications that allows the Calypso®Beacons to move in different directions to better simulate truelung-motion.Methods and Materials.A Calypso lung QA-phantom was designed, and 3D-printed. The design consists of three independent arms where the transponders were attached. A pinpoint-chamber with a buildup-cap was also incorporated. A 4-axis robotic arm was programmed to drive the motion-phantom to mimic breathing. After acquiring a four-dimensional-computed-tomography (4DCT) scan of the motion-phantom, treatment-plans were generated and delivered on a Varian TrueBeam®with Calypso capabilities. Stationary and gated-treatment plans were generated and delivered to determine the dosimetric difference between gated and non-gated treatments. Portal cine-images were acquired to determine the temporal-accuracy of delivery by calculating the difference between the observed versus expected transponders locations with the known speed of the transponders' motion.Results.Dosimetric accuracy is better than the TG142 tolerance of 2%. Temporal accuracy is greater than, TG142 tolerance of 100 ms for beam-on, but less than 100 ms for beam-hold.Conclusions.The robotic QA-phantom designed and developed in this study provides an independent phantom for performing Calypso lung-QA for commissioning and acceptance testing of Calypso for lung treatments.
Keywords: 3D printing; Calypso; motion management; quality assurance; robotics.
© 2021 Institute of Physics and Engineering in Medicine.
Similar articles
-
Commissioning and routine quality assurance of the Radixact Synchrony system.Med Phys. 2022 Feb;49(2):1181-1195. doi: 10.1002/mp.15410. Epub 2022 Jan 6. Med Phys. 2022. PMID: 34914846
-
A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy.J Appl Clin Med Phys. 2020 Jul;21(7):16-28. doi: 10.1002/acm2.12876. Epub 2020 Apr 13. J Appl Clin Med Phys. 2020. PMID: 32281265 Free PMC article.
-
Preliminary clinical experience with Calypso anchored beacons for tumor tracking in lung SBRT.Med Phys. 2020 Sep;47(9):4407-4415. doi: 10.1002/mp.14300. Epub 2020 Jun 25. Med Phys. 2020. PMID: 32463925
-
Advances in anthropomorphic thorax phantoms for radiotherapy: a review.Biomed Phys Eng Express. 2022 Jul 13;8(5). doi: 10.1088/2057-1976/ac369c. Biomed Phys Eng Express. 2022. PMID: 34736235 Review.
-
3D Motion Manipulation for Micro- and Nanomachines: Progress and Future Directions.Adv Mater. 2024 Jan;36(1):e2305925. doi: 10.1002/adma.202305925. Epub 2023 Nov 27. Adv Mater. 2024. PMID: 37801654 Review.
Cited by
-
Shaping success: clinical implementation of a 3D-printed electron cutout program in external beam radiation therapy.Front Oncol. 2023 Aug 9;13:1237037. doi: 10.3389/fonc.2023.1237037. eCollection 2023. Front Oncol. 2023. PMID: 37621682 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous