Use of an Extra-Tympanic Membrane Electrode to Record Cochlear Microphonics with Click, Tone Burst and Chirp Stimuli

Audiol Res. 2021 Mar 1;11(1):89-99. doi: 10.3390/audiolres11010010.

Abstract

This study determined electrocochleography (ECochG) parameter settings to obtain cochlear microphonics (CM) with less invasive flexible extra-tympanic membrane electrodes. In 24 adult normal-hearing subjects, CMs were elicited by presenting click stimuli at 100 dBnHL, tone bursts (2 kHz) and broadband (BB) CE-chirps® LS (Interacoustics, Middelfart, Denmark), both at 80 dBnHL. Different high-pass filters (HPFs) (3.3 Hz and 100 Hz, respectively) were used to investigate response quality of the CM. CMs were successfully obtained in 92-100% with click-, 75-83% with 2 kHz tone burst- and 58-63% with CE-chirp®-LS stimuli. Click stimuli elicited significantly larger CM amplitudes compared to 2 kHz tone bursts and BB CE-chirp® LS (Interacoustics, Middelfart, Denmark). No significant differences were found between the two different high-pass filter (HPF) settings. The present study shows that it is possible to obtain clear CMs with the flexible extra-tympanic membrane electrodes using click stimuli. In contrast to 2 kHz tone bursts and CE-chirp® (Interacoustics, Middelfart, Denmark) LS, clicks show a significantly higher success rate and are the preferred stimuli to confirm the presence or absence of CMs.

Keywords: auditory evoked potentials; cochlear microphonic; electrocochleography; tympanic membrane electrode.