Validation of the Thermo Scientific™ SARS-CoV-2 RT-PCR Detection Workflow for the Detection of SARS-CoV-2 from Stainless-Steel Environmental Surface Swabs: AOAC Performance Tested MethodSM 012103
- PMID: 33822958
- PMCID: PMC8083558
- DOI: 10.1093/jaoacint/qsab050
Validation of the Thermo Scientific™ SARS-CoV-2 RT-PCR Detection Workflow for the Detection of SARS-CoV-2 from Stainless-Steel Environmental Surface Swabs: AOAC Performance Tested MethodSM 012103
Abstract
Background: The Thermo Scientific™ SARS-CoV-2 reverse transcription-polymerase chain reaction (RT-PCR) Detection Workflow, packaged with Applied Biosystems™ TaqMan™ 2019-nCoV Assay Kit v1 targets three different SARS-CoV-2 genomic regions in a single RT-PCR reaction.
Objective: To validate the Thermo Scientific SARS-CoV-2 RT-PCR Workflow, for the detection of SARS-CoV-2 virus on stainless-steel surfaces as part of the AOAC Performance Tested MethodSM Emergency Response Validation program.
Method: The Applied Biosystems TaqMan 2019-nCoV Assay Kit v1, as part of the Thermo Scientific SARS-CoV-2 RT-PCR Workflow, was evaluated for specificity using in silico analysis of 15 764 SARS-CoV-2 sequences and 65 exclusivity organisms. The Thermo Scientific SARS-CoV-2 RT-PCR Workflow was evaluated in an unpaired study for one environmental surface (stainless steel) and compared to the U.S. Centers for Disease Control and Prevention 2019-Novel Coronavirus RT-PCR Diagnostic Panel, Instructions for Use (Revision 4, Effective 6/12/2020).
Results: In silico analysis showed that, of the 15 756 target SARS-CoV-2 genomes analyzed, 99% of the strains/isolates are perfectly matched to at least two of the three assays, and more than 90% have 100% homology to all three assays (ORF1ab, N-gene, S-gene) in the SARS-CoV-2 Kit. None of the 65 non-target strain genomes analyzed showed matching sequences. In the matrix study, the Thermo Scientific SARS-CoV-2 workflow showed comparable detection to the centers of disease control and prevention (CDC) method.
Conclusions: The Thermo Scientific SARS-CoV-2 RT-PCR Workflow is an effective procedure for detection of RNA from SARS-CoV-2 virus from stainless steel.
Highlights: The workflow provides equivalent performance results with the two tested RNA extraction platforms and the two tested RT-PCR instruments.
© AOAC INTERNATIONAL 2021. All rights reserved. For permissions, please email: journals.permissions@oup.com.
Similar articles
-
Validation Study for VERIPRO® SARS-CoV-2 Env Assay for the Detection of SARS-CoV-2 from Stainless-Steel Environmental Surface Swabs: Emergency Response Validation-AOAC Performance Tested MethodSM 122001.J AOAC Int. 2021 Aug 20;104(4):914-923. doi: 10.1093/jaoacint/qsab048. J AOAC Int. 2021. PMID: 33871628 Free PMC article.
-
Validation Study for PathogenDx EnviroX-Rv Assay for the Detection of SARS-CoV-2 from Stainless Steel Environmental Surface Swabs Performance Tested MethodSM 122003.J AOAC Int. 2021 Aug 20;104(4):889-913. doi: 10.1093/jaoacint/qsab047. J AOAC Int. 2021. PMID: 33839772 Free PMC article.
-
Detection of SARS-CoV-2 Virus on Stainless-Steel Surfaces: AOAC Performance Tested MethodSM 022102.J AOAC Int. 2021 Aug 20;104(4):924-934. doi: 10.1093/jaoacint/qsab049. J AOAC Int. 2021. PMID: 34416762 Free PMC article.
-
Detection methods for SARS-CoV-2 in tissue.Pathologe. 2021 Nov;42(Suppl 1):81-88. doi: 10.1007/s00292-021-00920-1. Epub 2021 Mar 26. Pathologe. 2021. PMID: 33770236 Free PMC article. Review.
-
[Methods of SARS-CoV-2 detection in tissue].Pathologe. 2021 Mar;42(2):208-215. doi: 10.1007/s00292-021-00919-8. Epub 2021 Mar 1. Pathologe. 2021. PMID: 33646360 Free PMC article. Review. German.
Cited by
-
Antisense oligonucleotides targeting ORF1b block replication of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).Front Microbiol. 2022 Oct 26;13:915202. doi: 10.3389/fmicb.2022.915202. eCollection 2022. Front Microbiol. 2022. PMID: 36386681 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical