Complications and Revision Rates in Minimally Invasive Robotic-Guided Versus Fluoroscopic-Guided Spinal Fusions: The MIS ReFRESH Prospective Comparative Study

Spine (Phila Pa 1976). 2021 Dec 1;46(23):1661-1668. doi: 10.1097/BRS.0000000000004048.

Abstract

Study design: Prospective, multicenter, partially randomized.

Objective: Assess rates of complications, revision surgery, and radiation between Mazor robotic-guidance (RG) and fluoro-guidance (FG).

Summary of background data: Minimally invasive surgery MIS ReFRESH is the first study designed to compare RG and FG techniques in adult minimally invasive surgery (MIS) lumbar fusions.

Methods: Primary endpoints were analyzed at 1 year follow-up. Analysis of variables through Cox logistic regression and a Kaplan-Meier Survival Curve of surgical complications.

Results: Nine sites enrolled 485 patients: 374 (RG arm) and 111 (FG arm). 93.2% of patients had more than 1 year f/u. There were no differences for sex, Charlson Comorbidity Index, diabetes, or tumor. Mean age of RG patients was 59.0 versus 62.5 for FG (P = 0.009) and body mass index (BMI) was 31.2 versus 28.1 (P< 0.001). Percentage of smokers was almost double in the RG (15.2% vs. 7.2%, P = 0.029). Surgical time was similar (skin-to-skin time/no. of screws) at 24.9 minutes RG and 22.9 FG (P = 0.550). Fluoroscopy during surgery/no. of screws was 15.5 seconds RG versus 35.4 seconds FG, (15 seconds average reduction). Fluoroscopy time during instrumentation/no. of screws was 3.6 seconds RG versus 17.8 seconds FG showing an 80% average reduction of fluoro time/screw in RG (P < 0.001). Within 1 year follow-up, there were 39 (10.4%) surgical complications RG versus 39 (35.1%) FG, and 8 (2.1%) revisions RG versus 7 (6.3%) FG. Cox regression analysis including age, sex, BMI, CCI, and no. of screws, demonstrated that the hazard ratio (HR) for complication was 5.8 times higher FG versus RG (95% CI: 3.5-9.6, P < 0.001). HR for revision surgery was 11.0 times higher FG versus RG cases (95% CI 2.9-41.2, P < 0.001).

Conclusion: Mazor robotic-guidance was found to have a 5.8 times lower risk of a surgical complication and 11.0 times lower risk for revision surgery. Surgical time was similar between groups and robotic-guidance reduced fluoro time per screw by 80% (approximately 1 min/case).Level of Evidence: 2.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Fluoroscopy
  • Humans
  • Lumbar Vertebrae
  • Minimally Invasive Surgical Procedures
  • Pedicle Screws*
  • Prospective Studies
  • Retrospective Studies
  • Robotic Surgical Procedures* / adverse effects
  • Spinal Fusion* / adverse effects