Abstract
Background: The ADHEAR™ system (MED-EL, Innsbruck, Austria) is a nonsurgical bone conduction device (BCD) to treat conductive hearing loss (CHL) and single-sided deafness. In contrast to the nonsurgical alternatives on headbands or spectacle frames, the audio processor of ADHEAR is placed retroauricularly on an adhesive adapter. The published evidence on the performance of this system is limited to studies with a trial period of 2-8 weeks.
Objective: This study assesses audiological and subjective outcomes over a period of 12 months, on patients with congenital aural atresia (CAA) using the ADHEAR hearing system.
Method: Fifteen children (mean age: 9.4 ± 4 years; range: 5-16 years) diagnosed with CAA (7 uni/8 bilateral) were included in this prospective, observational, repeated-measures study. Each subject used ADHEAR for 1 year, and the performance was evaluated after 1, 6, and 12 months. Free-field audiometry and speech discrimination tests were performed, and hearing-, general health- and device-specific questionnaires were used.
Results: The unaided sound field threshold improved from an average PTA4 of 63.6 ± 3.4 dB HL to an aided average PTA4 of 29.3 ± 3.0 dB HL after 1 month of device use. The word recognition score (WRS) improved from an average of 27.9 ± 15.9% unaided to an aided average WRS of 91.3 ± 4.4% (p = 0.0003) after 1 month, 92.0 ± 4.1% (p = 0.0002) after 6 months, and 92.7 ± 5.3% (p < 0.0001) after 12 months using the ADHEAR system compared to the unaided condition for all 3 time points. The improvements in the speech in noise at 1, 6, and 12 months were as well consistent over time. The average improvement at the signal to noise ratio (SNR) of +5 dB was 58% and 53% at the SNR of +0 dB. No complications were reported, and all patients continued to use the ADHEAR after the study end. The questionnaire results revealed high user satisfaction and an average wearing time of 12 h per day.
Conclusion: This 12-month trial of the nonsurgical adhesive BCD in CAA patients showed sufficient and reliable audiological and subjective outcomes, long wearing time, and high acceptance. The ADHEAR can be considered a suitable option to treat children with CAA for the given indication, without the drawbacks of nonsurgical devices that use pressure for retention of the audio processor or the costs and possible complications involved with a surgical alternative.
Keywords: Adhesive hearing aid; Aural atresia; Bone conduction; Conductive hearing loss; Microtia.
© 2021 S. Karger AG, Basel.
Similar articles
-
Audiological benefit and subjective satisfaction of children with the ADHEAR audio processor and adhesive adapter.Int J Pediatr Otorhinolaryngol. 2020 Feb;129:109729. doi: 10.1016/j.ijporl.2019.109729. Epub 2019 Oct 15. Int J Pediatr Otorhinolaryngol. 2020. PMID: 31689608
-
Auditory Performance and Subjective Satisfaction with the ADHEAR System.Audiol Neurootol. 2021;26(1):1-10. doi: 10.1159/000507775. Epub 2020 Jun 16. Audiol Neurootol. 2021. PMID: 32544908
-
A new adhesive bone conduction hearing system effectively treats conductive hearing loss in children.Int J Pediatr Otorhinolaryngol. 2019 Jul;122:117-125. doi: 10.1016/j.ijporl.2019.03.014. Epub 2019 Apr 3. Int J Pediatr Otorhinolaryngol. 2019. PMID: 31004837
-
Multicentric study on surgical information and early safety and performance results with the Bonebridge BCI 602: an active transcutaneous bone conduction hearing implant.Eur Arch Otorhinolaryngol. 2023 Apr;280(4):1565-1579. doi: 10.1007/s00405-022-07792-y. Epub 2023 Jan 10. Eur Arch Otorhinolaryngol. 2023. PMID: 36625869 Free PMC article. Review.
-
Long-term follow-up and review of the Bone Conduction Implant.Hear Res. 2022 Aug;421:108503. doi: 10.1016/j.heares.2022.108503. Epub 2022 Apr 3. Hear Res. 2022. PMID: 35490057 Review.
Cited by
-
[Nonsurgical attachment possibilities for bone conduction hearing aids : Comparison of coupling using external pressure versus an adhesive plate].HNO. 2023 May 16. doi: 10.1007/s00106-023-01300-2. Online ahead of print. HNO. 2023. PMID: 37191925 German.
-
Improvement in quality of life comparing noninvasive versus invasive hearing rehabilitation in children.Laryngoscope Investig Otolaryngol. 2023 Feb 23;8(2):591-598. doi: 10.1002/lio2.1030. eCollection 2023 Apr. Laryngoscope Investig Otolaryngol. 2023. PMID: 37090862 Free PMC article.
-
An objective bone conduction verification tool using a piezoelectric thin-film force transducer.Front Neurosci. 2022 Nov 17;16:1068682. doi: 10.3389/fnins.2022.1068682. eCollection 2022. Front Neurosci. 2022. PMID: 36466173 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous