Introduction: The EVOLVE implant (Wright Medical Technology, Arlington, TN, USA) is a modular loose-fitting radial head prosthesis. The primary objective was to synthesize all available literature investigating the midterm clinical outcomes of the EVOLVE implant.
Materials and methods: An electronic literature search in Pubmed/Medline, Scopus, EMBASE, and Cochrane library was performed querying for studies published in 2000-2017. Articles describing clinical and radiographical outcomes as well as reoperation were included. Outcomes of interest included range of motion, Mayo Elbow Performance Score, Disabilities of the Arm Shoulder and Hand, radiographic outcome, and reason for reoperation.
Results: A total of five articles consisting of 146 patients with EVOLVE implants were included. Mean patient age was 57.4 years (range 22-84), and 43.8% were males (n = 64). Mean follow-up was 4.8 years (range 1-14). Mean Mayo Elbow Performance Score and Disabilities of the Arm Shoulder and Hand score were 87.6 (range 30-100) and 18.9 (range 0-82), respectively. Midterm clinical results were good or excellent (Mayo Elbow Performance Score > 74) in 94 patients. Reoperation was observed in 12 patients, with implant revision required in 2 patients. The primary reason for reoperation was persistent stiffness (n = 9).
Conclusion: Midterm outcomes of EVOLVE radial head prosthesis are satisfactory, and associated complication rates are low. Loose-fit implant method appears to be a reliable approach to avoid failure of radial head prosthesis by painful loosening.
Keywords: anchorage; arthroplasty; failure; loose-fit; outcome; radial head prosthesis.