Study on the intrinsic detection efficiency of scintillator/Cherenkov detector for monitoring 14MeV neutrons by using foil activation method

Appl Radiat Isot. 2021 Aug:174:109761. doi: 10.1016/j.apradiso.2021.109761. Epub 2021 May 3.

Abstract

A neutron detector, based on scintillator and Cherenkov detector, was designed to monitor the D-T neutron generator in a PGNAA online measurement system in our previous study. In this paper, the foil activation method was used to study the intrinsic detection efficiency of the detector to the D-T neutron generator. The Fe foil with 99.99% purity was selected as the activation foil. The experiments and the GEANT4 simulations were carried out to study the intrinsic detection efficiency of the monitor. The results show that the intrinsic detection efficiency of the monitor is 20.42% ± 0.76%. Then, the calibrated detector and Fe foils were simultaneously used to measure the neutron flux of D-T neutron generator. The relative deviation between the results of the two methods is 5.64%.

Keywords: Cherenkov detector; Foil activation method; Intrinsic detection efficiency; Neutron monitor; Scintillator detector.