Disassociation of an Anteriorly Implanted Dual-Mobility Component with Migration to the Distal Posterior Thigh: A Case Report

JBJS Case Connect. 2021 May 17;11(2). doi: 10.2106/JBJS.CC.20.00400.

Abstract

Case: Intraprosthetic dislocation, a rare complication of modular dual-mobility hip replacements, occurs when the polyethylene component becomes dislodged. Our patient sustained an intraprosthetic dislocation with subsequent implant migration to the posterior thigh. In an anterior approach revision surgery at an outside institution, the component was unable to be located. Subsequent imaging revealed marked implant migration. Computed tomography (CT)-guided needle localization was used to facilitate component removal.

Conclusion: In the presence of different surgical approaches, dissociated polyethylene components may migrate to anatomic compartments distinct from the approach of implantation. CT and needle localization may be used to facilitate component removal.

Publication types

  • Case Reports

MeSH terms

  • Hip Dislocation* / surgery
  • Hip Prosthesis* / adverse effects
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Thigh / surgery