Aim: Patient-reported pad-count as continence rate assessment tool after artificial urinary sphincter (AUS) implantation is common. However, lack of standardized continence definition using this method results in heterogeneous published efficacy outcomes. Data on 24-h pad weight tests (PWT) after primary AUS implantation for postprostatectomy urinary incontinence (PPUI) is scarce. Our aim was to evaluate the 24-h PWT as an efficacy assessment tool and correlate it to qualitative outcomes using validated questionnaires.
Methods: This retrospective, single center, follow-up cohort study, evaluated 180 patients who underwent primary AUS implantation for PPUI from 2005 to 2018. Voiding diaries, 24-h PWT, validated patient satisfaction and quality of life (QoL) questionnaires were collected pre-operatively and at 3-6 months postactivation, using the institution's Electronic Medical Records.
Results: The median preoperative and postoperative 24-h PWT values were 494 (interquartile range [IQR]: 304-780) and 7 (IQR: 0-25) g respectively with a significant improvement in urinary leakage of 489.5 g 99.1% (p < 0.001). Median preoperative and postoperative I-QoL results increased from 33.5 (IQR: 19.3-63.6) to 86.4 (IQR: 73.9-94.3) points, with a significant 52.9 points improvement in QoL (p < 0.001). Similarly, the median preoperative and postoperative ICIQ-SF values decreased from 20 (IQR: 17-21) to 5 (IQR: 3-9) points, showing a significant improvement of 15 points (p < 0.001). We also found a significant correlation between PWT and patient satisfaction.
Conclusion: The 24-h PWT provides a reliable and objective assessment of continence rates, with a strong correlation to qualitative outcomes, after primary AUS implantation for PPUI. Its use could help reduce reported outcome heterogeneity across studies.
Keywords: artificial urinary sphincter; pad test; postprostatectomy urinary incontinence; quality of life; urinary incontinence.
© 2021 The Authors. Neurourology and Urodynamics published by Wiley Periodicals LLC.