In vivo kinematic comparison between an ultra-congruent and a posterior-stabilized total knee arthroplasty design by RSA
- PMID: 34117894
- DOI: 10.1007/s00167-021-06629-1
In vivo kinematic comparison between an ultra-congruent and a posterior-stabilized total knee arthroplasty design by RSA
Abstract
Purpose: The aim of the present study was to compare the in vivo under weight-bearing kinematic behavior of a posterior-stabilized (PS) and an ultra-congruent (UC) total knee arthroplasty (TKA) model during a sit-to-stand motor task, a common activity of daily life.
Methods: A cohort of 16 randomly selected patients (8 PS Persona Zimmer, 8 UC Persona Zimmer) was evaluated through dynamic radiostereometric analysis (RSA) at a minimum of 9 months after TKA, during the execution of a sit-to-stand. The anteroposterior (AP) translation of the femoral component and the AP translation of the low point of medial and lateral femoral compartments were compared through Student's t test (p < 0.05).
Results: A significantly greater anterior translation of the femoral component was found for the PS group compared to the UC group. The flexion interval where statistical significance was found was between 30° and 0° (p = 0.017). Both groups showed a significantly greater anterior translation of the low point of the lateral compartment with respect to the medial one (PS: p = 0.012, UC: p = 0.018). This was consistent with a medial-pivot pattern. Furthermore, a significantly greater anterior translation of the medial compartment was found in the PS group compared to the UC group (p = 0.001). The same pattern was observed for the lateral compartment (p = 0.006).
Conclusions: The TKA designs evaluated in the present study showed comparable in-vivo kinematics with regards to medial pivot pattern but differences in absolute AP translation. Specifically, the UC design showed greater AP stability than the PS design. This finding could be positive in terms of implant stability, but negative in terms of premature polyethylene wear and thus implant failure. This remains to be verified in studies with a larger sample size and longer follow-up.
Level of evidence: IV.
Keywords: Kinematics analysis; Posterior-stabilized design; Radiostereometric analysis; TKA; Ultra-congruent design.
© 2021. European Society of Sports Traumatology, Knee Surgery, Arthroscopy (ESSKA).
Similar articles
-
No kinematical difference between ultra-congruent and medial-congruent total knee arthroplasty when implanted with mechanical alignment: an in vivo dynamic RSA study.Knee Surg Sports Traumatol Arthrosc. 2022 Sep;30(9):2975-2979. doi: 10.1007/s00167-022-07033-z. Epub 2022 Jun 29. Knee Surg Sports Traumatol Arthrosc. 2022. PMID: 35768651 Free PMC article.
-
J-curve design total knee arthroplasty: the posterior stabilized shows wider medial pivot compared to the cruciate retaining during chair raising.Knee Surg Sports Traumatol Arthrosc. 2020 Sep;28(9):2883-2892. doi: 10.1007/s00167-019-05645-6. Epub 2019 Aug 3. Knee Surg Sports Traumatol Arthrosc. 2020. PMID: 31377823
-
Wider translations and rotations in posterior-stabilised mobile-bearing total knee arthroplasty compared to fixed-bearing both implanted with mechanical alignment: a dynamic RSA study.Knee Surg Sports Traumatol Arthrosc. 2023 Nov;31(11):4969-4976. doi: 10.1007/s00167-023-07541-6. Epub 2023 Aug 24. Knee Surg Sports Traumatol Arthrosc. 2023. PMID: 37615718 Free PMC article.
-
Comparison between gaits after a medial pivot and posterior stabilized primary total knee arthroplasty: a systematic review of the literature.Arthroplasty. 2023 Mar 17;5(1):15. doi: 10.1186/s42836-023-00165-8. Arthroplasty. 2023. PMID: 36927464 Free PMC article. Review.
-
The Medial Pivot Design in Total Knee Arthroplasty.Orthop Clin North Am. 2024 Jan;55(1):49-59. doi: 10.1016/j.ocl.2023.06.007. Epub 2023 Aug 5. Orthop Clin North Am. 2024. PMID: 37980103 Review.
Cited by
-
Mobile-Bearing Total Ankle Replacement In Vivo Kinematic Assessment: A Prospective Study Protocol.J Clin Med. 2022 Sep 10;11(18):5328. doi: 10.3390/jcm11185328. J Clin Med. 2022. PMID: 36142975 Free PMC article.
-
Native knee kinematics is not reproduced after sensor guided cruciates substituting total knee arthroplasty.J Exp Orthop. 2023 Feb 14;10(1):17. doi: 10.1186/s40634-023-00567-2. J Exp Orthop. 2023. PMID: 36786878 Free PMC article.
-
Favorable outcomes of posterior cruciate-stabilized inserts over ultracongruent inserts in total knee arthroplasty.Arch Orthop Trauma Surg. 2023 Sep;143(9):5857-5865. doi: 10.1007/s00402-023-04837-7. Epub 2023 Mar 16. Arch Orthop Trauma Surg. 2023. PMID: 36928502
-
No kinematical difference between ultra-congruent and medial-congruent total knee arthroplasty when implanted with mechanical alignment: an in vivo dynamic RSA study.Knee Surg Sports Traumatol Arthrosc. 2022 Sep;30(9):2975-2979. doi: 10.1007/s00167-022-07033-z. Epub 2022 Jun 29. Knee Surg Sports Traumatol Arthrosc. 2022. PMID: 35768651 Free PMC article.
-
A prospective comparison of total knee arthroplasty using ultra-congruent, condylar-stabilizing, and posterior-stabilized devices implanted with kinematic alignment: better 2-year outcomes with ultra-congruent.Knee Surg Sports Traumatol Arthrosc. 2023 Mar;31(3):1026-1033. doi: 10.1007/s00167-022-07206-w. Epub 2022 Nov 1. Knee Surg Sports Traumatol Arthrosc. 2023. PMID: 36318322 Free PMC article. Clinical Trial.
References
-
- Cates HE, Komistek RD, Mahfouz MR, Schmidt MA, Anderle M (2008) In vivo comparison of knee kinematics for subjects having either a posterior stabilized or cruciate retaining high-flexion total knee arthroplasty. J Arthroplasty 23:1057–1067 - DOI
-
- Walker PS, Sussman-Fort JM, Yildirim G, Boyer J (2009) Design features of total knees for achieving normal knee motion characteristics. J Arthroplasty 24(3):475–483 - DOI
-
- Kim TW, Lee SM, Seong SC, Lee S, Jang J, Lee MC (2016) Different intraoperative kinematics with comparable clinical outcomes of ultracongruent and posterior stabilized mobile-bearing total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24(9):3036–3043 - DOI
-
- Laskin RS, Maruyama Y, Villaneuva M, Bourne R (2000) Deep-dish congruent tibial component use in total knee arthroplasty: a randomized prospective study. Clin Orthop Relat Res 380:36–44 - DOI
-
- Akti S, Karakus D, Sezgin EA, Cankaya D (2020) No differences in clinical outcomes or isokinetic performance between cruciate-substituting ultra-congruent and posterior stabilized total knee arthroplasties: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-020-06275-z (Epub ahead of print) - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials