Abstract

Objectives: To evaluate the performance of 4-dimensional computed tomography (4D-CT) in assessing upper airway obstruction (UAO) in patients with Robin sequence (RS) and compare the accuracy and reliability of 4D-CT and flexible fiber-optic laryngoscopy (FFL).

Study design: Prospective survey of retrospective clinical data.

Setting: Single, tertiary care pediatric hospital.

Methods: At initial and 30-day time points, a multidisciplinary group of 11 clinicians who treat RS rated UAO severity in 32 sets of 4D-CT visualizations and FFL videos (dynamic modalities) and static CT images. Raters assessed UAO at the velopharynx and oropharynx (1 = none to 5 = complete) and noted confidence levels of each rating. Intraclass correlation and Krippendorff alpha were used to assess intra- and interrater reliability, respectively. Accuracy was assessed by comparing clinician ratings with quantitative percentage constriction (QPC) ratings, calculated based on 4D-CT airway cross-sectional area. Results were compared using Wilcoxon rank-sum and signed-rank tests.

Results: There was similar intrarater agreement (moderate to substantial) with 4D-CT and FFL, and both demonstrated fair interrater agreement. Both modalities underestimated UAO severity, although 4D-CT ratings were significantly more accurate, as determined by QPC similarity, than FFL (-1.06 and -1.46 vs QPC ratings, P = .004). Overall confidence levels were similar for 4D-CT and FFL, but other specialists were significantly less confident in FFL ratings than were otolaryngologists (2.25 and 3.92, P < .0001).

Conclusion: Although 4D-CT may be more accurate in assessing the degree of UAO in patients with RS, 4D-CT and FFL assessments demonstrate similar reliability. Additionally, 4D-CT may be interpreted with greater confidence by nonotolaryngologists who care for these patients.

Keywords: 4D-CT; Robin sequence; computed tomography; flexible fiber-optic laryngoscopy; upper airway obstruction.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, N.I.H., Extramural

MeSH terms

  • Child
  • Four-Dimensional Computed Tomography
  • Humans
  • Laryngoscopy* / methods
  • Pierre Robin Syndrome* / diagnostic imaging
  • Prospective Studies
  • Reproducibility of Results
  • Retrospective Studies