Mass recovery of individual polypeptides may be estimated under various practical conditions. With the purpose of obtaining rapid and reliable standard procedures for recovery measurements, we have compared five individual methods utilizing a silica-based stationary phase [Nucleosil C18 (7 microns)/ammonium sulphate-perchlorate-acetonitrile, pH 3.0] and a resin-based stationary phase (TSK Phenyl 5 PW RP/ammonium phosphate-acetonitrile, pH 7.0). The recoveries of insulin (6 kilodaltons), human growth hormone (22 kilodaltons) and human serum albumin (68 kilodaltons) estimated under five different experimental conditions were found to be concordant. Variations in column load, flow-rate, gradient shape and column dwell time and addition of cyclame did not increase the (reduced) recovery of serum albumin and growth hormone.