Objectives: Resonance frequency analysis (RFA) is used to monitor implant stability. Its output, the Implant Stability Quotient (ISQ), supposedly correlates with insertion torque, a common measurement of primary stability. However, the reliability of RFA in condensed bone remains unclear.
Material and methods: In this human cadaver study in edentulous jaws and fresh extraction sockets, implants were inserted using a split-mouth approach into condensed or untreated bone. Mean ISQ, peak insertion torque, and pre- and postoperative bone volume fractions (BV/TV) were assessed.
Results: In edentulous jaws, insertion torque and ISQ correlated both in untreated (r = 0.63, p = 0.02) and in condensed (r = 0.82, p < 0.01) bone. In extraction sockets, insertion torque and ISQ only correlated in untreated (r = 0.78, p < 0.01), but not in condensed bone (r = 0.15, p = 0.58). In all edentulous jaws, preoperative BV/TV correlated with insertion torque (r = 0.90, p < 0.0001), ISQ (r = 0.64, p < 0.001), and changes in BV/TV (r = -0.71, p < 0.01). In all extraction sockets, preoperative BV/TV did not correlate with either insertion torque (r = 0.33, p = 0.15), ISQ (r = 0.38, p = 0.09), or changes in BV/TV (r = -0.41, p = 0.09). Joint analysis identified preoperative BV/TV as a predictor of postoperative BV/TV (p < 0.001), insertion torque (p < 0.001), and ISQ (p < 0.001).
Conclusions: RFA is feasible for monitoring stability after late implant placement into condensed bone, but not after immediate placement into condensed fresh extraction sites.
Keywords: bone condensation; edentulous jaw; immediate implant placement; implant stability; resonance frequency analysis.
© 2021 The Authors. Clinical Oral Implants Research published by John Wiley & Sons Ltd.