Intraoperative Fracture of a Dual Modular Delta Ceramic Femoral Head During Total Hip Arthroplasty: A Case Report

JBJS Case Connect. 2021 Aug 16;11(3). doi: 10.2106/JBJS.CC.21.00215.

Abstract

Case: A 39-year-old man with cerebral palsy and hip dysplasia status post right Chiari osteotomy presented with right hip osteoarthritis in consultation for total hip arthroplasty (THA). During THA, a Delta ceramic head was misaligned on the taper, but this was obscured by an overlying dual modular polyethylene shell. When we attempted to tap the head onto the taper, fracture occurred requiring revision to a cobalt-chromium head due to slight taper damage.

Conclusion: To our knowledge, this is the first reported case of intraoperative fracture of a BIOLOX Delta dual modular head.

Publication types

  • Case Reports

MeSH terms

  • Adult
  • Arthroplasty, Replacement, Hip*
  • Ceramics
  • Femur Head
  • Hip Prosthesis* / adverse effects
  • Humans
  • Male
  • Prosthesis Design
  • Prosthesis Failure