Modelling optically pumped magnetometer interference in MEG as a spatially homogeneous magnetic field
- PMID: 34418526
- DOI: 10.1016/j.neuroimage.2021.118484
Modelling optically pumped magnetometer interference in MEG as a spatially homogeneous magnetic field
Abstract
Here we propose that much of the magnetic interference observed when using optically pumped magnetometers for MEG experiments can be modeled as a spatially homogeneous magnetic field. We show that this approximation reduces sensor level variance and substantially improves statistical power. This model does not require knowledge of the underlying neuroanatomy nor the sensor positions. It only needs information about the sensor orientation. Due to the model's low rank there is little risk of removing substantial neural signal. However, we provide a framework to assess this risk for any sensor number, design or subject neuroanatomy. We find that the risk of unintentionally removing neural signal is reduced when multi-axis recordings are performed. We validated the method using a binaural auditory evoked response paradigm and demonstrated that removing the homogeneous magnetic field increases sensor level SNR by a factor of 3. Considering the model's simplicity and efficacy, we suggest that this homogeneous field correction can be a powerful preprocessing step for arrays of optically pumped magnetometers.
Copyright © 2021. Published by Elsevier Inc.
Similar articles
-
Precision magnetic field modelling and control for wearable magnetoencephalography.Neuroimage. 2021 Nov 1;241:118401. doi: 10.1016/j.neuroimage.2021.118401. Epub 2021 Jul 15. Neuroimage. 2021. PMID: 34273527 Free PMC article.
-
Transforming and comparing data between standard SQUID and OPM-MEG systems.PLoS One. 2022 Jan 19;17(1):e0262669. doi: 10.1371/journal.pone.0262669. eCollection 2022. PLoS One. 2022. PMID: 35045107 Free PMC article.
-
Data-driven model optimization for optically pumped magnetometer sensor arrays.Hum Brain Mapp. 2019 Oct 15;40(15):4357-4369. doi: 10.1002/hbm.24707. Epub 2019 Jul 11. Hum Brain Mapp. 2019. PMID: 31294909 Free PMC article.
-
Wireless optically pumped magnetometer MEG.Neuroimage. 2024 Oct 15;300:120864. doi: 10.1016/j.neuroimage.2024.120864. Epub 2024 Sep 24. Neuroimage. 2024. PMID: 39322096
-
A 20-channel magnetoencephalography system based on optically pumped magnetometers.Phys Med Biol. 2017 Nov 10;62(23):8909-8923. doi: 10.1088/1361-6560/aa93d1. Phys Med Biol. 2017. PMID: 29035875 Free PMC article.
Cited by
-
Measurement of Frontal Midline Theta Oscillations using OPM-MEG.Neuroimage. 2023 May 1;271:120024. doi: 10.1016/j.neuroimage.2023.120024. Epub 2023 Mar 12. Neuroimage. 2023. PMID: 36918138 Free PMC article.
-
Using OPMs to measure neural activity in standing, mobile participants.Neuroimage. 2021 Dec 1;244:118604. doi: 10.1016/j.neuroimage.2021.118604. Epub 2021 Sep 21. Neuroimage. 2021. PMID: 34555493 Free PMC article.
-
Applications of OPM-MEG for translational neuroscience: a perspective.Transl Psychiatry. 2024 Aug 24;14(1):341. doi: 10.1038/s41398-024-03047-y. Transl Psychiatry. 2024. PMID: 39181883 Free PMC article. Review.
-
Using optically pumped magnetometers to replicate task-related responses in next generation magnetoencephalography.Sci Rep. 2024 Mar 18;14(1):6513. doi: 10.1038/s41598-024-56878-6. Sci Rep. 2024. PMID: 38499615 Free PMC article.
-
Spatial and Frequency Specific Artifact Reduction in Optically Pumped Magnetometer Recordings.J Integr Neurosci. 2022 Aug 16;21(5):145. doi: 10.31083/j.jin2105145. J Integr Neurosci. 2022. PMID: 36137963 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources