Comparison of Allplex™ SARS-CoV-2 Assay, Easy SARS-CoV-2 WE and Lumipulse quantitative SARS-CoV-2 antigen test performance using automated systems for the diagnosis of COVID-19

Int J Infect Dis. 2021 Dec:113:113-115. doi: 10.1016/j.ijid.2021.09.069. Epub 2021 Sep 29.

Abstract

Diagnostic methods based on SARS-CoV-2 antigen detection are a promising alternative to SARS-CoV-2 RNA amplification. We evaluated the automated chemiluminescence-based Lumipulse® G SARS-CoV-2 Ag assay as compared to real time assays (combined results from RT-PCR Allplex™ SARS-CoV-2 assay and Easy SARS-CoV-2 WE kit) on 513 nasopharyngeal swabs (NPS). Among these, 53.6% resulted positive to RT-PCR, considered as the reference test. Compared to the reference test, overall sensitivity and specificity of Lumipulse® G SARS-CoV-2 Ag assay were 84.0%, and 89.1%, respectively, and overall agreement between the antigen and molecular assays was substantial (κ = 0.727). When stratifying samples into groups based on ranges of RT-PCR cycle threshold (Ct), the antigen test sensitivity was >95% for samples with Ct <30. Linear regression analysis showed strong and highly significant correlation between the Lumipulse Ag concentrations and the RT-PCR Ct values (RdRp gene), irrespective of whether the Ct values from molecular test were combined in a unique regression analysis or analysed separately. Overall, chemiluminescence-based antigen assay may be reliably applied to NPS samples to identify individuals with high viral loads, more likely to transmit the virus.

Keywords: Antigen assay; COVID-19; Lumipulse; RT-PCR; SARS-CoV-2.

MeSH terms

  • COVID-19*
  • Humans
  • Nasopharynx
  • RNA, Viral / genetics
  • SARS-CoV-2*
  • Sensitivity and Specificity
  • Viral Load

Substances

  • RNA, Viral