Dual energy CT arthrography in shoulder instability: successful iodine removal with virtual non-contrast images and accurate 3D reformats of the glenoid for assessment of bone loss

Skeletal Radiol. 2022 May;51(5):1027-1036. doi: 10.1007/s00256-021-03916-3. Epub 2021 Oct 3.

Abstract

Objective: To evaluate the image quality of dual energy CT (DECT) of the shoulder after arthrography and of virtual non-contrast (VNC) 3D reformats of the glenoid and to compare glenoid measurements on VNC 3D reformats and on 2D CTs.

Materials and methods: DECT arthrography (80 kV/140 kV) was performed in 42 shoulders of 41 patients with instability using diluted iodinated contrast media (80 mg/ml). VNC images and VNC 3D reformats of the glenoid were calculated using image postprocessing. Dose parameters, CT values of intraarticular iodine and muscle, image contrast (iodine/muscle), and image quality (5-point scale: 1 = worst, 5 = best) were evaluated. Two independent readers assessed glenoid morphology and performed glenoid measurements on 2D and 3D images.

Results: Calculation of VNC images and VNC 3D reformats was successful in 42/42 shoulders (100%). The effective dose was mean 1.95 mSv (± 0.9 mSv). CT values of iodine and muscle were mean 1014.6 HU (± 235.8 HU) and 64.5 HU(± 8.6 HU), respectively, and image contrast was mean 950.2 HU (± 235.5 HU). Quality of cross-sectional images, VNC images, and VNC 3D reformats was rated good (median 4 (4-5), 4 (3-4), 4 (3-5), respectively). Detection of an osseous defect was equal on 2D and 3D images (13/42, P > 0.99) with no difference for measurement of the glenoid diameter with mean 28.3 mm (± 2.8 mm) vs. 28.4 mm (± 2.9 mm) (P = 0.5), width of the glenoid defect with 3.2 mm (± 2.1 mm) vs. 3.1 mm (± 2.3 mm) (P = 0.84), surface area with 638.5 mm2 (± 127 mm2) vs. 640.8 mm2 (± 129.5 mm2) (P = 0.47), and surface area of the defect with 46.6 mm2 (± 44.3 mm2) vs. 47.2 mm2 (± 48.0 mm2) (P = 0.73), respectively.

Conclusion: DECT shoulder arthrography is feasible and allows successful iodine removal with generation of VNC images and accurate VNC 3D reformats of the glenoid for assessment of bone loss.

Keywords: Arthrography; Computed tomography, X-ray; Image enhancement; Shoulder.