The VitMin Lab Sandwich-ELISA Assays for Iron and Inflammation Markers Compared Well with Clinical Analyzer Reference-Type Assays in Subsamples of the Nepal National Micronutrient Status Survey

J Nutr. 2022 Jan 11;152(1):350-359. doi: 10.1093/jn/nxab355.

Abstract

Background: The low cost and small specimen volume of the VitMin Lab ELISA assays for serum ferritin (Fer), soluble transferrin receptor (sTfR), C-reactive protein (CRP), and α-1-acid glycoprotein (AGP) have allowed their application to micronutrient surveys conducted in low-resource countries for ∼2 decades.

Objectives: We conducted a comparison between the ELISA and reference-type assays used in the US NHANES.

Methods: Using the Roche clinical analyzer as a reference, we measured random subsets of the 2016 Nepal National Micronutrient Status Survey (200 serum samples from children aged 6-59 mo; 100 serum samples from nonpregnant women) for Fer, sTfR, CRP, and AGP. We compared the combined data sets with the ELISA survey results using descriptive analyses.

Results: The Lin's concordance coefficients between the 2 assays were ≥0.89 except for sTfR (Lin's ρ = 0.58). The median relative difference to the reference was as follows: Fer, -8.5%; sTfR, 71.2%; CRP, -19.5%; and AGP, -8.2%. The percentage of VitMin samples agreeing within ±30% of the reference was as follows: Fer, 88.5%; sTfR, 1.70%; CRP, 74.9%; and AGP, 92.9%. The prevalence of abnormal results was comparable between the 2 assays for Fer, CRP, and AGP, and for sTfR after adjusting to the Roche assay. Continued biannual performance (2007-2019) of the VitMin assays in CDC's external quality assessment program (6 samples/y) demonstrated generally acceptable performance.

Conclusions: Using samples from the Nepal survey, the VitMin ELISA assays produced mostly comparable results to the Roche reference-type assays for Fer, CRP, and AGP. The lack of sTfR assay standardization to a common reference material explains the large systematic difference observed for sTfR, which could be corrected by an adjustment equation pending further validation. This snapshot comparison together with the long-term external quality assessment links the survey data generated by the VitMin Lab to the Roche assays used in NHANES.

Keywords: C-reactive protein; ferritin; inflammation; iron deficiency; micronutrients; soluble transferrin receptor; α-1-acid glycoprotein.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adolescent
  • Adult
  • Anemia, Iron-Deficiency* / diagnosis
  • Anemia, Iron-Deficiency* / epidemiology
  • Biomarkers
  • C-Reactive Protein / metabolism
  • Child
  • Enzyme-Linked Immunosorbent Assay
  • Female
  • Humans
  • Inflammation
  • Iron*
  • Micronutrients
  • Middle Aged
  • Nepal
  • Nutrition Surveys
  • Receptors, Transferrin
  • Young Adult

Substances

  • Biomarkers
  • Micronutrients
  • Receptors, Transferrin
  • C-Reactive Protein