Applicability, efficacy, and safety of over-the-scope clips in children

Gastrointest Endosc. 2022 Mar;95(3):489-499. doi: 10.1016/j.gie.2021.10.011. Epub 2021 Oct 16.

Abstract

Background and aims: Over-the-scope clips (OTSCs) are now becoming popular in endoscopy performed in adults for indications such as acute nonvariceal GI bleeding, anastomotic bleeding, and for closure of gastrocutaneous and postgastrostomy fistulae. Varied sizes of clip are available, but even the smallest, 8.5 to 9.8 mm in diameter with its loading device on the tip of the endoscope, increases device and endoscope intubation diameters up to 14.65 mm. This may present challenges in terms of the size of the patient in whom it might be used. OTSCs appear to be effective and safe in the hands of those who are trained appropriately in endoscopy on adult patients; however, the experience of OTSC application in children is not reported. Here we present results of a service evaluation of this technology at 2 regional/national referral pediatric endoscopy units in the United Kingdom and the United States.

Methods: Two tertiary centers' databases were searched to identify cases in which OTSCs were used. Demographics, presentation, anthropometry, comorbidities, efficacy, adverse events, and postprocedure follow-up were recorded, with identification of resolution or recurrence.

Results: OTSC procedures were performed on 24 occasions in 20 patients (11 girls) between February 2018 and February 2021. Patients had a mean age of 12 years (range, 5-17) and a mean weight of 44.42 kg (range, 18.2-70.3). Indications were nonhealing PEG site fistulae (n = 7), acute nonvariceal upper GI bleeding (ANUGIB) from gastric ulcers (5), ANUGIB from duodenal ulcers (3), nonhealing bleeding anastomotic ulcer (3), esophageal mucocutaneous fistula (1), and gastric perforation (1). Technical success was achieved in all but 1 case (95%), and clinical success was achieved in 18 cases (90%).

Conclusions: The OTSC device appears to be effective in children (minimum age 5 years and minimum weight 18 kg) in a limited number of situations including anastomotic ulcer, closure of leaking PEG site, gastric perforation, and bleeding peptic ulcers. The operator should be an experienced endotherapeutic endoscopist with specific OTSC training, and the type and size of the OTSC device should be carefully considered, along with any comorbidities of the patient that may preclude success and/or lead to potential adverse events such as esophageal perforation.

MeSH terms

  • Adult
  • Child
  • Child, Preschool
  • Endoscopy, Gastrointestinal / methods
  • Esophageal Fistula* / etiology
  • Female
  • Humans
  • Retrospective Studies
  • Stomach Ulcer* / etiology
  • Surgical Instruments
  • Treatment Outcome