A-scan ultrasound in ophthalmology: A simulation tool

Med Eng Phys. 2021 Nov:97:18-24. doi: 10.1016/j.medengphy.2021.09.005. Epub 2021 Sep 21.

Abstract

In the present study, we developed a computational tool for simulating the ophthalmological applications of A-scan ultrasound, including cataract characterisation and biometry. A-scan biometry is used to measure the axial length (AL) of the eye before cataract surgery to calculate the refractive power of the intraocular lens to be implanted. Errors in the measurement of the AL lead to post-surgical refractive errors. The simulation tool was developed using the k-Wave Matlab toolbox, together with a user-friendly interface developed in Matlab. Diverse error sources were evaluated. Constant ultrasound speed assumptions may introduce refractive errors of up to 0.6 D; by contrast, probe positioning errors had a lower impact, of up to 0.11 D. The correct identification of the Bruch's membrane is limited not only by axial resolution constraints but also by the low reflection coefficient at the retina/choroid interface. Regarding cataract characterisation, the amplitudes of the echoes reflected at the lens interfaces are sensitive to diverse cataract types and severities, and a more realistic representation could be obtained by using a higher resolution in the eye grid; however, the required computational times would make simulations impracticable when using personal computers. The simulation tool shows good versatility for evaluating diverse aspects of A-scan biometry.

Keywords: A-scan ultrasound; Biometry; Cataract; Eye axial length; Intraocular lenses; Simulation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biometry
  • Cataract Extraction*
  • Lenses, Intraocular*
  • Ophthalmology*
  • Refraction, Ocular