Background: Non-coplanar volumetric modulated arc therapy (ncVMAT) is proposed to reduce toxicity in heart and lungs for locoregional radiotherapy of left-sided breast cancer, including internal mammary nodes (IMN).
Patients and methods: This retrospective study included 10 patients with left-sided breast cancer who underwent locoregional radiotherapy after breast-conserving surgery. For each patient, the ncVMAT plan was designed with four partial arcs comprising two coplanar arcs and two non-coplanar arcs, with a couch rotating to 90°. The prescribed dose was normalized to cover 95% of planning target volume (PTV), with 50 Gy delivered in 25 fractions. For each ncVMAT plan, dosimetric parameters were compared with the coplanar volumetric modulated arc therapy (coV-MAT) plan.
Results: T here were improvements in conformity index, homogeneity index and V55 of total target volume (PTVall) comparing ncVMAT to coVMAT (p < 0.001). Among the organs at risk, the average V30, V20, V10, V5, and mean dose (Dmean) of the heart decreased significantly (p < 0.001). Furthermore, ncVMAT significantly reduced the mean V20, V10, V5, and Dmean of left lung and the mean V10 and V5 and Dmean of contralateral lung (p < 0.001). An improved sparing of the left anterior descending coronary artery and right breast were also observed with ncVMAT (p < 0.001).
Conclusions: Compared to coVMAT, ncVMAT provides improved conformity and homogeneity of whole P TV, better dose sparing of the heart, bilateral lungs, left anterior descending coronary artery (LAD), and right breast for locoregional radiotherapy of left-sided breast cancer with IMN, potentially reducing the risk of normal tissue damage.
Keywords: internal mammary nodes; left-sided breast cancer; modulated arc therapy; non-coplanar; volumetric.