Evaluation of a new smartphone optical blood pressure application (OptiBP™) in the post-anesthesia care unit: a method comparison study against the non-invasive automatic oscillometric brachial cuff as the reference method

J Clin Monit Comput. 2022 Oct;36(5):1525-1533. doi: 10.1007/s10877-021-00795-w. Epub 2022 Jan 3.

Abstract

We compared blood pressure (BP) values obtained with a new optical smartphone application (OptiBP™) with BP values obtained using a non-invasive automatic oscillometric brachial cuff (reference method) during the first 2 h of surveillance in a post-anesthesia care unit in patients after non-cardiac surgery. Three simultaneous BP measurements of both methods were recorded every 30 min over a 2-h period. The agreement between measurements was investigated using Bland-Altman and error grid analyses. We also evaluated the performance of the OptiBP™ using ISO81060-2:2018 standards which requires the mean of the differences ± standard deviation (SD) between both methods to be less than 5 mmHg ± 8 mmHg. Of 120 patients enrolled, 101 patients were included in the statistical analysis. The Bland-Altman analysis demonstrated a mean of the differences ± SD between the test and reference methods of + 1 mmHg ± 7 mmHg for mean arterial pressure (MAP), + 2 mmHg ± 11 mmHg for systolic arterial pressure (SAP), and + 1 mmHg ± 8 mmHg for diastolic arterial pressure (DAP). Error grid analysis showed that the proportions of measurement pairs in risk zones A to E were 90.3% (no risk), 9.7% (low risk), 0% (moderate risk), 0% (significant risk), 0% (dangerous risk) for MAP and 89.9%, 9.1%, 1%, 0%, 0% for SAP. We observed a good agreement between BP values obtained by the OptiBP™ system and BP values obtained with the reference method. The OptiBP™ system fulfilled the AAMI validation requirements for MAP and DAP and error grid analysis indicated that the vast majority of measurement pairs (≥ 99%) were in risk zones A and B.Trial Registration ClinicalTrials.gov Identifier: NCT04262323.

Keywords: Arterial hypertension; Hemodynamic; Hemodynamic monitoring; Mobile health; Mobile phone; Postoperative.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anesthesia*
  • Blood Pressure / physiology
  • Blood Pressure Determination / methods
  • Blood Pressure Monitors
  • Humans
  • Oscillometry
  • Smartphone*

Associated data

  • ClinicalTrials.gov/NCT04262323