Comparative Evaluation of Two NGAL Automated Immunoassays in Urine and Plasma

J Appl Lab Med. 2022 May 4;7(3):747-761. doi: 10.1093/jalm/jfab167.

Abstract

Background: Acute kidney injury (AKI), a frequent and serious complication of hospitalized patients, is associated with increased mortality and morbidity. Neutrophil gelatinase-associated lipocalin (NGAL) is a biomarker for the early identification of AKI. We report a comparative laboratory verification of the Abbott Diagnostics (ARCHITECT® urine NGAL) and BioPorto Diagnostics (NGAL TestTM) assays including an assessment of the Abbott assay's performance in EDTA plasma.

Methods: Intra-/interbatch imprecision, linearity, recovery, and limit of quantitation (LoQ) were assessed and an interassay comparison performed (n = 51). Between-laboratory agreement was assessed against other laboratories using the Abbott (n = 48) and BioPorto (n = 94) assays. Plasma NGAL (pNGAL) levels were measured in non-AKI patients with a range of estimated glomerular filtration rates (n = 80).

Results: Coefficients of variation (CVs) for intra- and interbatch imprecision were 0.7%-12.4% and 1.9%-27.5% for the BioPorto assay, respectively, and 1.4%-6.3%/3.4%-6.8%, respectively, for the Abbott assay. The BioPorto assay exhibited a higher LoQ (27.5 ng/mL vs 1.2 ng/mL). Both assays were linear over the range 5-6000 ng/mL. Recovery of recombinant NGAL was 113.1 ± 7.1% and 96.5 ± 7.8% for the Abbott and BioPorto assays, respectively. On average, the Abbott assay gave results 9.2% lower than the BioPorto assay. Mean differences of 0.2% (Abbott) and 20.2% (BioPorto) were observed in the between-laboratory comparison. In patients without AKI, pNGAL levels were inversely proportional to eGFR.

Conclusions: Performance of the Abbott and BioPorto assays was similar although the latter performed less well at lower NGAL concentrations. The Abbott assay tended to yield lower results, exhibited a lower LoQ and over-recovered NGAL. Although only Conformité Européenne-marked and marketed for use in urine, the Abbott assay demonstrated equivalent performance to the BioPorto assay with EDTA plasma.

Keywords: AKI; EDTA plasma; NGAL; immunoassay; urine; verification.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acute Kidney Injury* / diagnosis
  • Acute Kidney Injury* / urine
  • Acute-Phase Proteins / urine
  • Edetic Acid
  • Female
  • Humans
  • Immunoassay
  • Lipocalin-2
  • Lipocalins* / urine
  • Male
  • Proto-Oncogene Proteins / urine

Substances

  • Acute-Phase Proteins
  • Lipocalin-2
  • Lipocalins
  • Proto-Oncogene Proteins
  • Edetic Acid