Accelerometer-based portable navigation, a faster guide compared to computer-assisted navigation in bilateral total knee arthroplasty-a randomized controlled study

Knee Surg Sports Traumatol Arthrosc. 2023 Mar;31(3):786-792. doi: 10.1007/s00167-021-06842-y. Epub 2022 Jan 7.

Abstract

Purpose: There are limited studies in the literature comparing the alignment accuracy of the large console, imageless, computer-assisted navigation (CAN) and portable, hand-held, accelerometer-based navigation (ABN) in total knee arthroplasty (TKA). This study was aimed to compare the operative time, blood loss, radiological, clinical and functional outcomes between CAN- and ABN-guided bilateral TKA.

Methods: From Jan 2016 to Dec 2017, 50 patients who underwent bilateral TKA were randomized to undergo either CAN-guided or ABN-guided TKA. Tourniquet time and blood loss were recorded, and intra-op complications were noted. Post-op radiological outcomes at 2 weeks were compared between the groups. The clinical and functional outcomes using the American Knee Society Scores (KSS) and Oxford Knee Score (OKS) were recorded pre-operatively and post-operatively at 3, 6, 12 months and at the end of the study with a minimum follow-up of 48 months.

Results: Both groups were well-matched in terms of patient demographic parameters. The mean surgical time per knee was significantly lower in the ABN group (54.5 ± 5.6 min) compared to the CAN group (61.7 ± 13.7 min; p < 0.01). Mean blood loss per knee in the ABN group was 592.1 ± 245.3 mL compared to 682.8 ± 322.0 in the CAN group (p = 0.11). In the ABN group, the mean post-op mechanical axis was 1.2 ± 3.2° (vs 1.5 ± 2.2° in the CAN group, p 0.6), the frontal femoral angle was 88.8 ± 2.3° (vs 88.8 ± 1.8° in the CAN group, p 1.0) and frontal tibial angle was 90.1 ± 1.6° (vs 89.7 ± 1.1° in the CAN group, p 0.14). At 48 month follow-up, the mean functional KSS in the ABN group was 89.0 ± 5.7 (vs 88.1 ± 4.5 in the CAN group, p 0.37) and the mean OKS was 40.5 ± 2.8 (vs 39.6 ± 3.2 in the CAN group, p 0.12).

Conclusion: Portable, hand-held ABN offers alignment accuracy and functional outcomes in TKA similar to that with CAN, with a reduced duration of surgery. There was no advantage of either of the techniques in terms of clinical or functional outcomes at 48 month follow-up. LEVEL OF EVIDENCE 1.

Keywords: Accelerometer-based navigation; Computer navigation; Total knee arthroplasty.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Accelerometry / methods
  • Arthroplasty, Replacement, Knee* / methods
  • Computers
  • Humans
  • Knee Joint / surgery
  • Osteoarthritis, Knee* / surgery
  • Surgery, Computer-Assisted* / methods