Visual Acuity, Wavefront Aberrations, and Defocus Curves With an Enhanced Monofocal and a Monofocal Intraocular Lens: A Prospective, Randomized Study

J Refract Surg. 2022 Jan;38(1):10-20. doi: 10.3928/1081597X-20211109-02. Epub 2022 Jan 1.

Abstract

Purpose: To compare uniocular and binocular visual acuity, wavefront aberrations, and defocus curves using the TECNIS Eyhance (Johnson & Johnson) and RayOne (Rayner) intraocular lenses (IOLs).

Methods: In this prospective, randomized, comparative study, 50 patients (100 eyes) were randomized to receive the same IOL bilaterally (ClinicalTrials.gov Identifier: NCT04175951). Follow-up visits were at 1 and 3 to 9 months postoperatively. Primary outcome measures were uncorrected distance (UDVA) and uncorrected intermediate (UIVA) visual acuity (logMAR) at 66 cm. Secondary outcome measures were corrected distance (CDVA) and distance-corrected intermediate (DCIVA) visual acuity at 66 cm, manifest refraction, spherical (Z40) and vertical coma (Z3-1) aberrations (total, internal eye, and corneal) at normal pupil size, defocus curves, and Catquest 9SF and Glare and Halos questionnaire on Likert scale (1 = no glare and halos and 4 = continuous).

Results: Uniocular UDVA (P = .02), UIVA (P = .02), and binocular UIVA (P < .01) and uniocular (P = .01) and binocular (P < .01) DCIVA were better with the TECNIS Eyhance IOL at 3 to 9 months. At 3 to 9 months, both uniocular and binocular defocus curves were significantly broader with the TECNIS Eyhance IOL between -0.50 and -3.00 diopters (D). For total and internal eye, there was a statistically but clinically insignificant difference in Z40 with the TECNIS Eyhance IOL. Rasch scores improved (TECNIS Eyhance: 2.04 ± 1.34 to 2.91 ± 0.81 and RayOne:1.61 ± 1.35 to 2.97 ± 0.16) at 3 to 9 months. Refraction, Z3-1, glare, and halos were not different.

Conclusions: The TECNIS Eyhance IOL provided better DCIVA and broader defocus curves than the RayOne IOL. There was no difference in CDVA or patient-reported outcomes. Although there were some differences in aberrations when measured with normal pupil size, they were not clinically significant. [J Refract Surg. 2022;38(1):10-20.].

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Humans
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Patient Satisfaction
  • Phacoemulsification*
  • Prospective Studies
  • Prosthesis Design
  • Refraction, Ocular
  • Vision, Binocular
  • Visual Acuity

Associated data

  • ClinicalTrials.gov/NCT04175951