Evaluation of point-of-care thumb-size bispectral electroencephalography device to quantify delirium severity and predict mortality

Br J Psychiatry. 2021 Aug 2:1-8. doi: 10.1192/bjp.2021.101. Online ahead of print.

Abstract

Background: We have developed the bispectral electroencephalography (BSEEG) method for detection of delirium and prediction of poor outcomes.

Aims: To improve the BSEEG method by introducing a new EEG device.

Method: In a prospective cohort study, EEG data were obtained and BSEEG scores were calculated. BSEEG scores were filtered on the basis of standard deviation (s.d.) values to exclude signals with high noise. Both non-filtered and s.d.-filtered BSEEG scores were analysed. BSEEG scores were compared with the results of three delirium screening scales: the Confusion Assessment Method for the Intensive Care Unit (CAM-ICU), the Delirium Rating Scale-Revised-98 (DRS) and the Delirium Observation Screening Scale (DOSS). Additionally, the 365-day mortalities and the length of stay (LOS) in the hospital were analysed.

Results: We enrolled 279 elderly participants and obtained 620 BSEEG recordings; 142 participants were categorised as BSEEG-positive, reflecting slower EEG activity. BSEEG scores were higher in the CAM-ICU-positive group than in the CAM-ICU-negative group. There were significant correlations between BSEEG scores and scores on the DRS and the DOSS. The mortality rate of the BSEEG-positive group was significantly higher than that of the BSEEG-negative group. The LOS of the BSEEG-positive group was longer compared with that of the BSEEG-negative group. BSEEG scores after s.d. filtering showed stronger correlations with delirium screening scores and more significant prediction of mortality.

Conclusions: We confirmed the usefulness of the BSEEG method for detection of delirium and of delirium severity, and prediction of patient outcomes with a new EEG device.

Keywords: Delirium; bispectral EEG (BSEEG); delirium rating scales; electroencephalography; mortality.

Grants and funding