Investigating the effects of endogenous lipaemia on the measurement of sodium by indirect ion specific electrode potentiometry

Ann Clin Biochem. 2022 Sep;59(5):324-329. doi: 10.1177/00045632221098628. Epub 2022 May 5.

Abstract

Background: The widely automated method using indirect ion specific electrodes (ISE) potentiometry for determination of sodium concentration is prone to interference from lipaemia. Manufacturer-specified lipaemic (L)-index cut offs may underestimate the effects of endogenous lipaemia.

Methods: We assessed the interference on sodium concentration caused by endogenous lipaemia in 32 residual samples (from 13 patients) using indirect ISE (Cobas® 8000 modular analyser with c702 module, Roche diagnostics) and direct ISE (GEM 4000 premier, Werfen) potentiometric methods. Regression analysis (linear and non-linear) was used to determine a reliable (L)-index cut off for reporting sodium concentration.

Results: There was a poor correlation observed between triglyceride concentration and (L)-index. There was significant negative interference caused by endogenous lipaemia within analysed samples. Non-linear regression demonstrated a negative interference of approximately 5% at an (L)-index of 250.

Conclusion: At present, the manufacturer advises not to report sodium concentration by indirect ISE on the Cobas® 8000 modular analyser if the (L)-index is >2000. However, this has been determined by the addition of exogenous lipids (Intralipid®) and it is clear that this is not comparable to endogenous lipaemia. To ensure patient safety, clinical laboratories should consider lowering the cut off for (L)-index that they use for reporting sodium concentration.

Keywords: Lipaemia; interference; ion specific electrode (ISE); potentiometry; sodium.