Temperature-controlled high-power short-duration ablation with 90 W for 4 s: outcome, safety, biophysical characteristics and cranial MRI findings in patients undergoing pulmonary vein isolation

J Interv Card Electrophysiol. 2022 Nov;65(2):491-497. doi: 10.1007/s10840-022-01146-8. Epub 2022 Jun 24.

Abstract

Background: High-power short-duration (HPSD) radiofrequency ablation (RFA) is highly efficient and safe while reducing procedure and RF time in pulmonary vein isolation (PVI). The QDot™ catheter is a novel contact force ablation catheter that allows automated flow and power adjustments depending on the local tissue temperature to maintain a target temperature during 90 W/4 s lesions. We analysed intraprocedural data and periprocedural safety using the QDot-catheter in patients undergoing PVI for paroxysmal atrial fibrillation (PAF).

Methods: We included n = 48 patients undergoing PVI with the QDot-catheter with a temperature-controlled HPSD ablation mode with 90 W/4 s (TC-HPSD). If focal reconnection occurred besides repeat ablation, the ablation mode was changed to 50 W/15 s (QMode). N = 23 patients underwent cerebral MRI to detect silent cerebral lesions.

Results: Mean RF time was 8.1 ± 2.8 min, and procedure duration was 84.5 ± 30 min. The overall maximal measured catheter tip temperature was 52.0 °C ± 4.6 °C, mean overall applied current was 871 mA ± 44 mA and overall applied energy was 316 J ± 47 J. The mean local impedance drop was 12.1 ± 2.4 Ohms. During adenosine challenge, n = 14 (29%) patients showed dormant conduction. A total of n = 24 steam pops were detected in n = 18 patients (39.1%), while no pericardial tamponade occurred. No periprocedural thromboembolic complications occurred, while n = 4 patients (17.4%) showed silent cerebral lesion.

Conclusions: TC-HPSD ablation with 90 W/4 s using the QDot-catheter led to a reduction of procedure and RF time, while no major complications occurred. Despite optimized temperature control and power adjustment, steam pops occurred in a rather high number of patients, while none of them leads to tamponade or to clinical or neurological deficits.

Keywords: Ablation; Atrial fibrillation; High-power short duration; RF ablation.

MeSH terms

  • Atrial Fibrillation* / diagnostic imaging
  • Atrial Fibrillation* / etiology
  • Atrial Fibrillation* / surgery
  • Catheter Ablation* / methods
  • Equipment Design
  • Humans
  • Magnetic Resonance Imaging
  • Pulmonary Veins* / diagnostic imaging
  • Pulmonary Veins* / surgery
  • Steam
  • Temperature
  • Treatment Outcome

Substances

  • Steam