Impedance Planimetry (Endoflip) and Ideal Distensibility Ranges for Optimal Outcomes after Nissen and Toupet Fundoplication

J Am Coll Surg. 2022 Sep 1;235(3):420-429. doi: 10.1097/XCS.0000000000000273. Epub 2022 Aug 10.

Abstract

Background: Previous research has shown that impedance planimetry-based functional lumen imaging probe (FLIP) measurements are associated with patient-reported outcomes after laparoscopic antireflux surgery. We hypothesize that Nissen and Toupet fundoplications have different ideal FLIP profiles, such as distensibility.

Study design: A retrospective review of a prospectively maintained quality database was performed. Patients who had FLIP measurements during fundoplications between 2013 and 2021 were included. Reflux Symptom Index, Gastroesophageal Reflux Disease-Health Related Quality of Life Questionnaire, and dysphagia score were collected for up to 2 years postoperatively. The Wilcoxon rank-sum test was used to compare FLIP measurements vs outcomes.

Results: Two hundred fifty patients (171 Toupet, 79 Nissen) were analyzed. Distensibility ranges were categorized as tight, ideal, or loose. The ideal distensibility index range of Toupet patients with the 30- and 40-mL balloon fills were 2.6 to 3.7 mm2/mmHg. This range was associated with less dysphagia at 1 year compared with the tight group (p = 0.02). For Nissen patients, the 30- and 40-mL ideal threshold was a distensibility index of ≥2.2 mm2/mmHg. Patients with distensibility exceeding this threshold had a better quality of life than the tight group, reporting better Gastroesophageal Reflux Disease-Health Related Quality of Life Questionnaire (p = 0.02) and lower dysphagia scores (p = 0.01) at 2 years.

Conclusions: Impedance planimetry revealed different ideal distensibility ranges after Toupet and Nissen fundoplications that are associated with improved patient-reported outcomes, suggesting that intraoperative FLIP has the potential to tailor fundoplication.

MeSH terms

  • Deglutition Disorders* / diagnosis
  • Deglutition Disorders* / etiology
  • Electric Impedance
  • Fundoplication / methods
  • Gastroesophageal Reflux* / complications
  • Gastroesophageal Reflux* / diagnosis
  • Gastroesophageal Reflux* / surgery
  • Humans
  • Laparoscopy* / methods
  • Quality of Life
  • Treatment Outcome